• Require periodic electrical safety checks to ensure that the grounding system in your area
remains intact.
• If the transducer is left in a patient during periods when imaging is not taking place, unplug
the transducer from the system to reduce the possibility of leakage current or ESU
interaction. Also make sure that the deflection control brakes are off and that the
transducer is in freewheeling mode.
• Always remove the transducer from the patient before defibrillation.
TEE Deflection Control Basics
The deflection controls on the TEE transducer move the deflection area, located between the
distal tip and flexible shaft. The deflection area bends when you operate the controls,
permitting anterior, posterior, and lateral positioning.
Deflection Control Movement
To prevent tissue damage such as pressure necrosis, gastroesophageal lacerations, bleeding,
tearing of adhesions, ligament damage, and perforation, observe the following warnings. See
“TEE Transducer References” on page 238
WARNING
Never apply excessive force when inserting or withdrawing a TEE transducer, or when
operating the transducer deflection controls.
WARNING
Lock medial/lateral movement of the TEE transducer during insertion.
TEE Deflection Control Basics
Transesophageal Transducers
EPIQ 7 User Manual 4535 617 25341
207
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...