NOTE
When Hide Doppler Velocity Minus Sign is enabled, avoid using the Doppler velocity value
displayed in the results and report for manual calculations. The sign value, whether visible or
not, affects any calculations performed with the reported value. The system includes the sign
value in calculations.
To specify the display of the Doppler velocity minus sign:
1. Touch Utilities.
2. In the System tab, touch Setups.
3. Click Analysis.
4. Select the tab of the calculations package that you want to change, and then click Tools &
Results.
5. Under Hide Doppler Velocity Minus Sign, select On or Off.
6. Click Close.
System Options
In addition to the standard features available in the system, other features are available as
purchasable licensed options. The types of options available include clinical applications,
protocols, imaging capabilities, QLAB Advanced Quantification Software Q-Apps, PercuNav
Image Fusion and Interventional Navigation, and connectivity capabilities.
To add licensed options to your system, you purchase them from your Philips representative.
Once purchased, they are installed in your system by a Philips field service engineer.
Custom Procedures
You can automate a variety of workflows by creating custom procedures for specific patients,
transducers, and presets.
Creating a Custom Procedure
1. Touch Utilities.
2. On the System tab, touch Setups.
3. Click System Settings.
4. Click the Patient Data tab.
Customizing the System
System Options
152
EPIQ 7 User Manual 4535 617 25341
Содержание epiq 7
Страница 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Страница 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Страница 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Страница 100: ...Brake Steering Lock Pedal System Overview System Components 100 EPIQ 7 User Manual 4535 617 25341 ...
Страница 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Страница 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Страница 198: ...Intraoperative Transducers Leakage Current Testing for Intraoperative Transducers 198 EPIQ 7 User Manual 4535 617 25341 ...
Страница 244: ...Endocavity Transducers Biopsy with Endocavity Transducers 244 EPIQ 7 User Manual 4535 617 25341 ...
Страница 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...