vi
Efficia Rack Installation and Configuration Guide
Safety Standards
Parameter
Specification
ABNT NBR IEC 60601-1, ABNT NBR IEC 60601-1-2, ABNT NBR IEC 60601-1-1, ABNT NBR IEC
60601-1-4, ABNT NBR IEC 60601-2-26, ABNT NBR IEC 60601-2-27, ABNT NBR IEC 60601-2-30,
ABNT NBR IEC 60601-2-34, ABNT NBR IEC 60601-2-49
Protection Class
Class I, internally powered equipment, per ABNT NBR IEC
60601-1
Degree of Protection
Type CF defibrillator-proof: per ABNT NBR IEC 60601-1
IPX1 Ingress protection against vertically falling drops of water
Mode of Operation
Continuous
Protection Against Hazards of
Ignition
of
Flammable
Anaesthetic Mixtures
Equipment is not suitable for use in the presence of a flammable
anaesthetic mixture with air or oxygen or nitrous oxide, per ABNT
NBR IEC 60601-1
Safety
Anyone who connects additional equipment to the signal input port or signal output port configures a medical system
and is therefore responsible to ensure compliance with the system requirements of NBR IEC 60601-1. If in doubt,
contact the Philips Customer Care Solutions Center or your local Philips representative.
Warning
If multiple instruments are interconnected or if multiple instruments are connected to a patient, the sum of the
leakage currents may exceed the limits given in ABNT NBR IEC 60601-1 system requirements.
Electromagnetic interference may cause disruption of performance. Protect the monitor and the rack from
sources of intense electromagnetic radiation. This device is designed to provide resistance to electromagnetic
interference; however, because of the proliferation of radio-frequency transmitting equipment and other
sources of electrical noise (such as cellular phones, mobile two-way radios, and electrical appliances) in the
health care environment, it is possible that high levels of such interference due to close proximity or strength of
a source, may result in disruption of performance of this device. Disruption may be evidenced by erratic
readings, cessation of operation or other incorrect functioning. If this occurs, the site of use should be surveyed
to determine the source of this disruption, and actions taken to eliminate the source. If you need assistance,
contact the Philips Customer Care Solutions Center or your local Philips representative.