7 Testing and Maintenance
87
4
The expected test results depend on the relay contact used. Please check that the correct relay
activity is initiated during alarm condition.
Reporting of Test Results
Philips recommends all test results are documented in accordance with local laws. Authorized Philips
personnel report the test result back to Philips in a service record. While hospital personnel
(biomedical engineers or technicians) do not need to report results to Philips, Philips recommends that
they record and store the test results in accordance with local laws.
The following table lists what to record after completing the tests in this chapter. Record the results in
the empty column in the following table.
The following is a guide as to what your documentation should include:
•
Identification of the testing body (for example, which company or department carried out the
tests).
•
Name of the person(s) who performed the tests and the concluding evaluation.
•
Identification of the device(s) and accessories being tested (serial number, etc.).
•
The actual tests (incl. visual inspections, performance tests, safety, and system tests) and
measurements required.
•
Date of testing and of the concluding evaluation.
•
A record of the actual values of the test results, and whether these values passed or failed the tests.
•
Date and confirmation of the person who performed the tests and evaluation.
The device under test should be marked according to the test result: passed or failed.
Test
Expected test results
Multi-Port Nurse Call Connector Test
Correct relay activity is initiated during alarm
condition (pass/fail)
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...