Maintenance
WARNING
Schedule
: Failure on the part of the responsible individual hospital or institution employing the use of
this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure
and possible health hazards.
In case of problems
: If you discover a problem with any of the equipment, contact your service
personnel, Philips, or your authorized supplier.
Electric shock hazard
: Do not open the monitor housing. Refer all servicing to qualified service
personnel.
Inspecting the Equipment and Accessories
You should perform a visual inspection
before each use
, and in accordance with your hospital's
policy. With the monitor switched off:
1
Examine unit exteriors for cleanliness and general physical condition. Make sure that the housings
are not cracked or broken, that everything is present, that there are no spilled liquids that may have
entered the housing, and that there are no signs of abuse.
2
Inspect all accessories (transducers, sensors and cables, and so forth). Do not use a damaged
accessory.
3
Switch the monitor on and make sure the display is bright enough. If the brightness is not
adequate, contact your service personnel or your supplier.
Inspecting the Cables and Cords
1
Examine all system cables, the power plug and cord for damage. Make sure that the prongs of the
plug do not move in the casing. If damaged, replace it with an appropriate power cord.
2
Inspect the cables, leads and their strain reliefs for general condition. Make sure there are no
breaks in the insulation. Make sure that the connectors are properly engaged at each end to
prevent rotation or other strain.
3
Carry out performance assurance checks as described in the monitor's
Service Guide
.
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...