20
161
20
Printing the ECG Waveform
FM30/50
You can print the ECG wave onto the trace paper. If you are monitoring both DECG and MECG,
both waves will be printed. The start of the wave recording is annotated above the wave with
MECG
for Maternal ECG, with
DECG
for Direct fetal ECG, and with
25 mm/sec
below the wave.
WARNING
The fetal/maternal monitor is not a diagnostic ECG device. In particular, the display of fetal/maternal
ECG is intended only for evaluating signal quality for fetal/maternal heart rate as derived from the
ECG waveform.
When in doubt, it can be used to identify sources of compromised signal quality, such as noise or
muscle artifacts. It can subsequently be used to verify the result of measures taken to resolve them
(e.g., checking ECG cable connections or adapting the fetal
ArtifactSuppress
configuration).
The safety and effectiveness of the displayed fetal/maternal ECG waveform (i.e., P, QRS and T
segments) for evaluation of fetal/maternal cardiac status during labor have not been evaluated.
The ECG waveform is printed along the bottom of the heart rate grid, and the three different
possibilities look like this:
DECG waveform on its own
MECG waveform on its own
1
DECG
2
Paper speed
1
MECG
2
Paper speed
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...