LIST OF DOCUMENTATION IN THIS BINDER:

SUBSYSTEM MANUAL BuckyDiagnost FS IsoRAD

{

UNIT MANUAL manual/automatic collimator

Note:

indicates document present

Содержание 9890 010 83921

Страница 1: ...dical Systems ALL RIGHTS RESERVED 4512 984 28681 REV AA SYSTEM BUCKYDIAGNOST FS ISORAD LIST OF DOCUMENTATION IN THIS BINDER SUBSYSTEM MANUAL BuckyDiagnost FS IsoRAD UNIT MANUAL manual automatic collimator Note indicates document present ...

Страница 2: ...ENTS DRAWINGS SCHEMATIC DIAGRAMS WIRING DIAGRAMS 1 2 3 4 5 6 Z2 Z Z1 W Philips Medical Systems 4512 984 28681 REV AA 1 BD_FS_IsoRAD_reg SERVICE MANUAL 704 SUBSYSTEM BuckyDiagnost FS IsoRAD 9890 010 83921 DMC Hamburg Printed in Hamburg Germany E 2004 Philips Medical Systems ALL RIGHTS RESERVED ...

Страница 3: ... 0 1 1 5 04 0 _________________________________ 2 1 04 0 _________________________________ 3 1 04 0 _________________________________ 4 0 1 04 0 4 1 10 04 0 _________________________________ 5 1 04 0 _________________________________ 6 0 1 04 0 6 1 3 04 0 _________________________________ Z 0 1 04 0 Z 1 1 04 0 A3 Z 1 2 04 0 A3 Z 1 3 04 0 A4 Z 1 7 01 0 A4 4512 982 0078 2Z 10 b 04 0 A3 4512 983 0757...

Страница 4: ...ntroduction and technical data 1 1 1 1 Purpose of the manual 1 1 1 2 Items supplied 1 1 1 3 Compatibility 1 2 1 4 Technical data 1 2 1 4 1 Mechanical data 1 2 1 4 2 Environmental data 1 2 1 4 3 Electrical data 1 3 1 4 4 Tools Material required 1 3 1 5 Safety information 1 4 1 6 Compliance information 1 4 1 7 Equipment identification and labeling 1 4 1 8 Abbreviations and definitions 1 5 1 9 Manual...

Страница 5: ...nd The floor stand IsoRAD is part of the BuckyDiagnost IsoRAD system shelter version The FS IsoRAD has a special design and is not available for other systems of the BuckyDiagnost family 1 2 Items supplied FS IsoRAD Column with weight counter balanced holder for X ray tube assembly Long tube carrier arm extendable Control handle option Automatic collimation and cassette size sensing Rail system fl...

Страница 6: ...ith other subsystems Generator OPTIMUS RAD US BuckyDiagnost VE BuckyDiagnost TH2 1 4 Technical data 1 4 1 Mechanical data For information about mechanical data refer to chapter Z Drawings 1 4 2 Environmental data Operation Stock Transport Temperature in Celcius 0 40 40 75 Temperature in Celcius Hour N A N A Humidity in non condensing 5 95 5 95 Gradient in hour N A N A Vibrations Shock range in Hz ...

Страница 7: ...age surge N A Voltage sag N A Static frequency variation 50 or 60 Hz 1 Hz Dynamic frequeny variation N A Harmonic voltage distortion single total N A Neutral to ground voltage variation N A Neutral to ground volate impulse N A Nominal current standby 50 60 Hz 0 5 0 6 A maximum current 50 60 Hz 2 3 2 6 A Nominal frequency 50 Hz 60 Hz Heat emission standby N A in operation N A Operating voltages 230...

Страница 8: ...that this equipment be operated in accordance with the guidelines set down by the national council on radiation protection ___________________ WARNING The light source of the light pointer is a Class II laser Never look into thebeam directly With Class II lasers the eye is protected from brief random glances into the laser beam by theeyelid closure reflex Class II lasers may therefore be used with...

Страница 9: ...efinitions Abbreviation Explanation BIST Build In Self Test FS Floor Stand PMS Philips Medical Systems POST Power On Self Test PRD Product Reference Data RO Rotalix SID Source Image Distance SRO Super Rotalix TC Tool Code TF Table Fixed TH Table Height adjustable VE Vertical VR Vertical Radiography VT Vertical Tiltable 1 9 Manual history Date Version Name Reason of changes 16 08 04 1 0 M Bierstedt...

Страница 10: ...ATION BuckyDiagnost FS IsoRAD 04 0 2 1 E 2004 Philips Medical Systems ALL RIGHTS RESERVED 2_Installation_04_BW 1 Installation The FS IsoRAD is already installed and aligned in the shelter For replacement activities refer to section 4 ...

Страница 11: ...nost FS IsoRAD FAULTFINDING BuckyDiagnost FS IsoRAD 04 0 3 1 E 2004 Philips Medical Systems ALL RIGHTS RESERVED 3_Fault_Finding_BW 1 Fault finding Use manual SMCM BuckyDiagnost TH2 manual Code No 4512 984 2565x ...

Страница 12: ... the longitudinal brake 4 2 1 2 2 Replacement of the rear wheel 4 3 1 3 Long tube carrier arm extendable 4 4 1 3 1 Replacement of the complete arm 4 4 1 3 2 Replacement of the base plate installed micro switch 4 4 1 3 3 Replacement of the catch 4 5 1 3 4 Replacement of the magnetic unit 4 6 1 3 5 Replacement of the alpha switch plate 4 6 1 3 6 Replacement of the transverse brake 4 8 1 4 Control gr...

Страница 13: ...t_12NC_04 1 FS IsoRAD 1 1 Overview over the covers When performing the replacement procedures the following covers may be affected 1 Column front cover top 2 Column front cover bottom 3 Rear bottom cover 4 Cable support cover 5 Electronic rack cover 6 Column base rear cover 7 Column base side cover rear 8 Column top cover 1 2 3 4 5 6 7 8 ...

Страница 14: ...n base side cover rear 6 7 of the carriage S Remove the rail anchor 1 with an Allen key Note Remember the order of the distance plates S Remove the distance plates 2 S Loosen the clamp 3 S Disconnect the cable 4 S Remove the brake 5 S Install the new brake S Connect the cable 4 S Fix the clamp 3 S Adjust the brake S Install the rail anchor 1 S Ensure that the wheel runs free If not change the posi...

Страница 15: ...so that the rear wheel runs free S Fix the position for instance with a wooden bar 1 S Remove the screw M6 2 with an Allen key Note Keep in mind the order of the distance rings on the mandrel S Remove the mandrel 1 S Remove the wheel 2 S Place the distance rings in correct order and the new wheel on the mandrel S Insert the mandrel and fix it Note Ensure that the order of the distance plates of th...

Страница 16: ... Replacement of the base plate installed micro switch Code no 4512 133 0544 S Remove the cover at the bottom of the tube carrier arm S Remove the connector 1 S Loosen the four screws 1 and replace the base plate installed micro switch Ensure the correct function of the micro switch S Check the function of the system S Re install the cover Replacement of the complete tube carrier arm 4 h Replacemen...

Страница 17: ...2NC_04 1 3 3 Replacement of the catch Code no 4512 133 0452 S Remove the cover at the bottom of the tube carrier arm S Remove the base plate installed micro switch according to chapter 1 3 2 S Remove the screws 1 and replace the catch S Install the base plate installed micro switch according to chapter 1 3 2 Replacement of the catch 0 5 h 1 ...

Страница 18: ...r arm S Remove the base plate installed micro switch according to chapter 1 3 2 S Loosen the four screws 1 and replace the magnetic unit S Install the base plate installed micro switch according to chapter 1 3 2 1 3 5 Replacement of the alpha switch plate Code No 4512 13 1441 S Switch the system OFF S Turn the control handle clockwise to its end position S Remove the cover 1 Replacement of the mag...

Страница 19: ...ical Systems ALL RIGHTS RESERVED 4_Replacements_mit_12NC_04 Use a wire and pliers for easier handling of the tension spring S Unhook the spring at its top 1 S Remove the nut 1 S Remove the upper screw of the alpha switch plate 1 S Pull out the alpha switch plate and remove the wires 1 1 1 ...

Страница 20: ...hode of both diodes UEX4 08 3 to UFA S1 1 UEX4 09 4 to UFA S1 4 UEX4 04 2 to UFA S2 1 S Insert and fix the alpha switch plate S Hook the spring S Switch the system ON S Ensure the 0 90 and 90 positions at control handle according to table The LED light is on only in 0 90 and 90 positions S1 S2 degree ON ON 0 ON OFF 90 OFF ON 90 S Re install the cover 1 3 6 Replacement of the transverse brake Code ...

Страница 21: ...ed the tube carrier arm can fall down Secure the tube carrier arm e g by lifting the table S Remove the end stopper by loosening the screws 1 S Remove the screws 1 from the brake plate Note Keep in mind the positions of the cable S Remove the cable 2 S Replace the brake 3 at the brake plate S Solder the cable S Fix the cable at the tube carrier arm S Fix the end stopper S Check the function of the...

Страница 22: ...S IsoRAD 4_Replacements_mit_12NC_04 1 4 Control grip 1 4 1 Replacement of the PCB control handle CU2 Code no 4512 108 0718 Note After replacement of the CU2 proceed height adjustment Refer to chapter 10 2 Adjustment of the ceiling suspension height in manual X Scope Replacement of the PCB on the control handle 1 h ...

Страница 23: ...rogramming of the UZ field 4512 108 0738x S Open link W1 1 1 2 Software programmings with X Scope Refer to manual X Scope in SRM BuckyDiagnost FS IsoRAD Coding field V3 W1 W2 R6 V2 R4 R5 C1 V1 R3 R2 R1 H3 H2 H1 F3 F2 F1 X14 X13 1 X21 X22 X25 X12 X11 X10 X5 X4 X3 X2 X1 1 1 1 1 1 1 1 1 1 UZ 4 0 A slow 3 2 A slow 3 2 A slow 24V 24V R6 V2 R5 V1 C1 W1 W2 UZX7 5 UZX2 2 1 ...

Страница 24: ...IGHTS RESERVED Adjstment 6_inh 6 0 1 ADJUSTMENTS Contents TEXT Contents 1 0 1 1 Adjustments 6 1 1 1 Alignment check of the column 6 1 1 2 Centering the system 6 1 1 3 Alignment to the table 6 1 1 4 Laser alignment 6 2 1 4 1 Center laser alignment 6 2 1 4 2 SID laser alignment 6 3 1 4 3 Final work 6 3 ...

Страница 25: ...ering the system S Switch the system ON S Fix the table bucky into the system center position S Switch ON the simulation light The cross line from the collimator and the cross line of the grid should coincide If necessary position the catch plates 1 3 Alignment to the table Note Check the correct alignment of the table Refer to Subsystem manual BuckyDiagnost TH2 section Installation chapter 2 S Ch...

Страница 26: ... on α 0 position the LED 1 ON S Remove the control handle grip 2 on the rearside by loosening screws and nuts 1 4 1 Center laser alignment S Press button 1 The center laser and the diaphragm light are ON S Raise the table top up and down S Observe the center laser light on the table top If the beam wanders 2 rotate the control handle by loosening four screws 3 on the rear side S Tighten screws the...

Страница 27: ...e top into the prefered value of the SID 900 1200 mm S Press button 2 center laser and pulsing SID laser and diaphragm light are ON S Observe the center laser and the pulses SID laser on the table top If the beams are not merged loosen the two fixing screws 3 and adjust the SID laser with the eccenter screw 4 Tighten the two fixing screws 3 1 4 3 Final work S Install the removed control handle gri...

Страница 28: ...edical Systems ALL RIGHTS RESERVED drawings_Z 0_1 Drawings FS Standard mechanic BuckyDiagnost FS IsoRAD mechanical dimensions Z 1 1 BuckyDiagnost FS IsoRAD floor rails mechanical dimensions Z 1 2 Installation of floor rail and wall rail Z 1 3 Cable outlet Z 1 7 Labelling 2Z 10 ...

Страница 29: ...yDiagnost FS IsoRAD Z 1 1 04 0 C Philips Medical Systems BuckyDiagnost FS ISO RAD A3 04 09 15 Schr Mechanical dimensions 101 196 125 217 2080 150 telescope arm min 400 max 2000 760 room height shelter 2160 1212 995 1000 ...

Страница 30: ...500 500 500 500 500 500 500 500 Floor rail 4900mm BuckyDiagnost FS Iso RAD 16 90 73 4900 500 200 200 Z 1 2 04 0 C Philips Medical Systems BuckyDiagnost FS ISO RAD A3 04 09 07 Schr Mechanical dimensions ...

Страница 31: ... rope 4mm 10 50 5 3 Schr 2062 16 8 4 54 M8 c BuckyDiagnost FS ISO RAD Philips Medical Systems 04 0 04 09 07 A4 Installation of Z 1 3 BuckyDiagnost FS Iso RAD 90 73 floor rail and wall rail M8 54 2080 80 plumbline ...

Страница 32: ......

Страница 33: ...limator RALCO or FDA Address Collimator E 2Z 10 b 04 0 C Philips Medical Systems 4512 983 07571 A1 A3 04 05 26 Schr BuckyDiagnost FS Labelling Laser details Laser details UL CSA Logo Address Typen No Electr data IEC 60601 1 IEC 60601 2 32 Identifications label Stand FS Identification label Tube holder ass Labeling as indicated in survey of tubes B BuckyDiagnost FS ...

Страница 34: ...UAN 1 Z1 2 BuckyDiagnost FS IsoRAD with sensing system cabling Z1 3 BuckyDiagnost FS detailed schematic Z1 4 BuckyDiagnost family control handle and automatic collimator Nicol Z1 5 UZ Coding field 4512 108 07385 Z1 7 Wiring diagrams BuckyDiagnost FS IsoRAD Earthing diagram Z2 1 BuckyDiagnost FS IsoRAD Wiring diagram Z2 2 BuckyDiagnost FS CAN bus cable Z2 3 BuckyDiagnost family control handle with ...

Страница 35: ...n UAGS2 Switch Tomo working position z direction UAN1 Power supply UB Column stand UBB Brake longitudinl Y direction UBE Catch and SID position switches UE Control handle UL1 Power filter UQ Transverse Telescope arm UQB1 Stroke magnet for beta rotation UQBS1 Switch for beta rotation position UQB3 Brake for lateral movement x direction UQBS2 Position switch for lateral movement x direction US Tube ...

Страница 36: ...5 2 X25 2 4 X25 1 3 X13 1 X13 2 X13 3 X13 4 X10 3 X10 1 UL1 Mains filter 1 3 4 2 X12 3 X12 1 UBN3 Tomo 220V 5 3 2A slow 3 2A slow 4 0A slow UAN1 24V 12V 12V DC AC yw gn bl br X1 1 DC AC X1 2 X1 3 yw gn bl br X3 2 X3 3 X4 1 X4 2 X4 3 X4 4 X14 3 6 X14 4 X14 2 5 UE control handle 24V UAD Tracking 24V LA Collimator 24V 12V green green green 24V 24V 24V 12V UZ L N L N UZ X9 1 X9 2 ...

Страница 37: ...M9 CAN in CAN out Rot Alpha switch M6 B X101 Control Handle UE LA US Automatic Collimator X4 X3 B X102 B X103 D9 B1 BS2 BS1 B3 Rot Beta stroke magnet Switch Beta Switch X Lateral movement Brake X Lateral movement UQ Transverse arm M9 M2 X6 X3 E B Catch Y and SID CAN N1 Power Supply Tube assembly support A 230V 12V C X1 M2 X1 1 D9 X4 X8 M9 M2 BX1 230V Y Brake Y Brake B Potenti ometer FFD M3 D9 X22 ...

Страница 38: ...6 5V X1 14 P5 9 X11 14 S8 DATA6 X11 07 H6 5V X1 07 DATA2 X11 03 H3 5V X1 03 DATA4 X11 05 H5 5V X1 05 DATA3 X11 04 H4 5V X1 04 DATA0 X11 01 H1 5V X1 01 H2 DATA1 X11 02 X1 02 5V UES26 S25 UES25 X15 7 S22 S24 S23 S20 S21 brake S1 X4 8 ALPHA_2 S2 X4 4 ALPHA_1 X4 2 POS_ALPHA X4 9 24V BX102 5 X4 6 BETA_0 CLUTCH_ALPHA UZX21 19 UZX23 2 UZX23 1 POS_Z X1 19 S1 UQ X1 02 X1 03 X1 04 X1 05 POS_X1 POS_X0 POS_X2...

Страница 39: ...X1 23 X11 23 X11 17 X1 17 24V X1 18 24V X11 18 X11 10 CS_7 X1 10 X11 9 CS_6 X1 9 X11 8 DATA7 X1 8 X11 7 DATA6 X1 7 X11 6 DATA5 X1 6 X11 5 DATA4 X1 5 X11 4 DATA3 X1 4 X11 3 DATA2 X1 3 X11 2 DATA1 X1 2 S12 S1 S2 S3 S4 S5 S6 S7 S8 S9 D13 S10 BuckyDiagnost family Control handle field E1 fan T1 M M4 M ruler M3 M special filter M2 M shutter B shutter A M1 M LA2 LA3 LA4 controller RSTCAN CAN_H CAN_L 5V 1...

Страница 40: ...tems 04 0 Z1 7 Coding field 4512 108 0738 V3 W1 W2 R6 V2 R4 R5 C1 V1 R3 R2 R1 H3 H2 H1 F3 F2 F1 X14 X13 1 X21 X22 X23 X24 X25 X12 X11 X10 X9 X8 X7 X6 X5 X4 X3 X2 X1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 UZ 4 0 A slow 3 2 A slow 3 2 A slow 24V 24V R6 V2 R5 V1 C1 W1 W2 UZX7 5 UZX2 2 ...

Страница 41: ...74 Corrugated hose support Screw Corrugated hose X75 Screw Front bottom Screw X15 X14 UZ X11 Main Filter UAN1 X1 3 X12 rd Screw X11 X12 X13 Screw MEX PE H T generator UB Column Column stand X16 X78 Tube assembly UQ Cover rear right X79 Cover rear left X78 Screw Power Rack UA X11 LA XA Tube Transverse Telescope arm option Tube assembly support Cover rear column top UA LA UE XA US UB UBE UQ UZ UAN1 ...

Страница 42: ... Column S4 S2 S1 L1 UBAX1 7 UBAX1 1 UBAX1 8 UBES5 1 UBAX1 2 2 1 1 2 UZX3 1 BY1 UZX3 2 Y1 UEX4 5 UEX4 6 Supporting arm tube assembly UB Column stand UBAX1 4 UEX4 5 04 0 ALL RIGHTS RESERVED C 1997 Philips Medical Systems DMC A3 04 09 01 Schr Wiring diagram Control handle UAF Position Z FFD Tomo safety switch Control handle UAN1X4 Tomo longitudinal center switch Tomo safety end switch Power supply Co...

Страница 43: ... UFN1 CAN Out CAN In CAN Out CAN In SZ1X3 SX90 4512 130 5314x UADX4 UADX3 UFX3 4512 130 5324x 4512 130 5324x LAX11 LAX1 UEX10 UEX9 LA CAN In CAN Out Tracking UAD CAN In CAN Out UE CAN In CAN Out CAN Out CAN In SZ1X3 SX90 TH2 UAX1 1 UFX3 4512 130 5324x 4512 130 5324x 4512 131 8742x LAX11 LAX1 UEX10 UEX9 4512 130 5314x CAN Out CAN In SZ1X3 SX90 TH2 5 Sensing Tracking Tomo 4 Tomo 3 Sensing Tomo CAN b...

Страница 44: ...8 H19 H20 H11 H12 H6 H9 H10 S7 S6 S5 S4 S3 S11 S9 S12 S10 S2 S1 H5 H4 S8 UE1 Control handle display UE2X11 X1 H2 H1 H7 H3 UE1X1 UADX4 UADX3 UBN2X3 UZX25 UFA UFB UZX7 Test Bar Code X13 X11 X12 X14 X3 W1 H1 X9 X10 X1 X15 S24 1 S25 1 S20 1 S20 4 S22 1 S21 1 S23 1 3 5 4 9 8 1 2 7 6 1 X4 1 X2 1 1 X5 X15 7 S26 4 S25 2 4 1 X15 6 S25 4 S24 4 S26 2 4 1 X15 2 X15 1 S21 4 S20 2 4 1 X15 3 S20 4 S22 4 S21 2 4 ...