Receiver VCXO Test
Chapter 8 - RF Bench Repair
4. Switch on the Receiver. The green LED on the Receiver Assembly
should flicker. Check the +5V, -12V and +12V voltages at the marked
capacitors.
Figure 22 Receiver Assembly Test
5. If all voltages are present and the green LED is flickering now test
Receiver VCXO.
6. If all voltages are correct and the green LED is not flickering, replace
the Receiver Assembly. Check for processor part number higher than
or equal to 1821-1141 on the Receiver Assembly. If any voltage is
incorrect replace the flat cable. If the problem is still not solved, replace
Receiver Processor Board (503).
Receiver VCXO Test
You require a Spectrum Analyzer and Active Probe to carry out this test.
1. Check the Receiver VCXO ouput signal at the transistor amplifier.
Содержание 50 T Series
Страница 4: ......
Страница 10: ...x List of Figures ...
Страница 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Страница 35: ...Cardio Channel Chapter 4 Troubleshooting Cardio Channel Figure 7 Troubleshooting Cardio Channel ...
Страница 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Страница 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Страница 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Страница 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Страница 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Страница 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Страница 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Страница 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...
Страница 102: ...USA Chapter 11 Channels and Frequencies ...