Fuses
Chapter 7 - Replacing Parts
Fuses
The fuse values are printed beside the mains socket:
For 100-120V
Line Voltage T300mA 250V
UL
For 220-240V
Line Voltage T125 L 250V
IEC
(
= alternating current)
For part numbers see the Receiver parts list.
To replace the fuses:
1. Switch off the Receiver and disconnect it from the main power supply.
2. Using a flat-blade screwdriver, turn the fuse cover (A) anti-clockwise.
3. Pull out the fuse holder and fuse.
4. Remove the fuse from the holder and replace it with another of the
correct value.
5. Slide the holder and fuse back into place.
6. Using a flat bladed screwdriver press in the fuse cover and
simultaneously turn the cover clockwise.
7. Repeat steps 2 and 6 for the second fuse.
Содержание 50 T Series
Страница 4: ......
Страница 10: ...x List of Figures ...
Страница 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Страница 35: ...Cardio Channel Chapter 4 Troubleshooting Cardio Channel Figure 7 Troubleshooting Cardio Channel ...
Страница 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Страница 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Страница 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Страница 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Страница 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Страница 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Страница 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Страница 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...
Страница 102: ...USA Chapter 11 Channels and Frequencies ...