Using the Service Program
Chapter 6 - Using the Service Software
Program S/N to
Transmitter
Programs the Serial Number to the Transmitter (if no Serial Number is
present on the Transmitter). This feature is needed when the Transmitter
Printed Circuit Board is exchanged.
Follow the steps as the program requests.
When moving the service connector from one device to the other, a "SIO
RCV error" may be displayed. This is due to the PC’s serial interface circuitry
and can be ignored.
If a Serial Number needs to be RESET, contact the factory Technical
Marketing Engineer to get RESET-CODE needed to RESET the Serial
Number.
Program S/N to
Receiver
Programs the Serial Number to the Receiver (if no Serial Number is present
on the Receiver). This feature is needed when Receiver-PC-Bd. is exchanged.
Follow the steps as the program requests.
When moving the service connector from one device to the other, a "SIO
RCV error" may be displayed. This is due to the PC’s serial interface circuitry
and can be ignored.
If a Serial Number needs to be RESET, contact factory Technical Marketing
Engineer to get RESET-CODE to RESET the Serial Number.
Power On Selftest
The PC acts as a Terminal. The Transmitter or Receiver connected to the PC
must be switched OFF and then ON. The PC displays the Start-up messages
from the connected Transmitter or Receiver.
Show last errors/
warnings
Shows the summary of the received Errors and Warnings.
Check Transmitter
Provides a Transmitter check of the following:
•
detected modes
•
status bits
•
TOCO value
Check Receiver
Provides a Receiver check of the following:
•
detected modes
•
status bits
•
TOCO value
•
RF field strength
Read SerNum and
Revisions
When selected from the Main Menu, this sub-menudisplays:
•
Software Revision
•
Serial Protocol Revision
•
Serial Number (S/N)
Содержание 50 T Series
Страница 4: ......
Страница 10: ...x List of Figures ...
Страница 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Страница 35: ...Cardio Channel Chapter 4 Troubleshooting Cardio Channel Figure 7 Troubleshooting Cardio Channel ...
Страница 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Страница 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Страница 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Страница 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Страница 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Страница 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Страница 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Страница 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...
Страница 102: ...USA Chapter 11 Channels and Frequencies ...