background image

 

 

 

Model No. UTG-2900 

 

 

 

 

 

 

 

 

 

283 Veterans Blvd 

Carlstadt, NJ. 07072 

(201) 933-6300 

www.phase2plus.com 

 

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание UTG-2900

Страница 1: ...Model No UTG 2900 283 Veterans Blvd Carlstadt NJ 07072 201 933 6300 www phase2plus com www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...9 4 2 Probe Zero 9 4 3 Sound Velocity Calibration 10 4 4 Making Measurements 12 4 5 Scan mode 13 4 6 Changing Resolution 13 4 7 Changing Units 13 4 8 Memory Management 14 4 9 Data Printing 15 4 10 Bee...

Страница 3: ...ring Range Pulse Echo 0 025 23 0 in steel Echo Echo 0 118 2 35 Accuracy 0 5 thickness 001 Sound velocity range 3280 32805 ft s 1000 9999m s Upper Lower Limit Preset Alarm Yes Scanning Feature Yes 16 m...

Страница 4: ...ickness of a part or structure by accurately measuring the time required for a short ultrasonic pulse generated by a transducer to travel through the thickness of the material reflect from the back or...

Страница 5: ...Indicates the coupling status While the gauge is taking a measurement the coupling indicator will be displayed If it is not the gauge is having difficulty achieving a stable measurement and the thick...

Страница 6: ...2 Save 3 Enter 4 Back Light 5 Probe Zero 6 Inch metric 7 Thru Coating on off 8 Scan on off 9 Calibration Velocity setting 10 Calibration block 0 157 4mm 1 5 6 7 3 2 8 9 4 10 www GlobalTestSupply com F...

Страница 7: ...f the material through which the sound travels has any grain structure the sound waves will experience scattering Both of these effects reduce the strength of the waves and thus the gauge s ability to...

Страница 8: ...ing measured and the other semicircle is responsible for conducting the echoed sound back into the transducer When the transducer is placed against the material being measured it is the area directly...

Страница 9: ...Transducers should be inspected on a regular basis for signs of uneven wear of the wearface If the wearface is worn on one side more than another the sound beam penetrating the test material may no l...

Страница 10: ...r factor and will compensate for this value in any subsequent measurements When performing a probe zero the instrument will always use the sound velocity value of the built in probe disc even if some...

Страница 11: ...on mode and return to the measurement mode The gauge is now ready to perform measurements 4 3 2 Calibration to a known velocity Note This procedure requires that the operator knows the sound velocity...

Страница 12: ...locities to enable the user to quickly check different known materials Example If you are testing steel at 233 in us you can quickly change to one of your stored sound velocities when you need to test...

Страница 13: ...gauge excels at making single point measurements it is sometimes desirable to examine a larger region searching for the thinnest point The gauge includes a feature called Scan Mode which allows it to...

Страница 14: ...the key to activate the data logging functions It will display the current file name and the total record count of the file 2 Use the key and the key to scroll to the file that will be cleared of all...

Страница 15: ...background light at any moment as you need after power on Since the EL light will consume more power turn on it only when necessary 4 12 Battery Icon Two AA size alkaline batteries are used as the pow...

Страница 16: ...f repairs must be supplied as well ANY ATTEMPT AT HOME REPAIR WILL AUTOMATICALLY VOID THE STATED WARRANTY NO EXCEPTIONS 6 Transport and Storage Be sure to clean the probe and cable after each use Grea...

Страница 17: ...ommended that the probe be left in contact with the surface for as short a time as needed to acquire a stable measurement While the transducer is in contact with a hot surface it will begin to heat up...

Страница 18: ...thus may be undesirable Other suitable couplants for measurements at normal temperatures may include water various oils and greases gels and silicone fluids Measurements at elevated temperatures will...

Страница 19: ...85 2200 Magnesium 0 230 5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0...