Model
No.
UTG
‐
4000
www.phase2plus.com
283 Veterans Blvd
Carlstadt, NJ. 07072
(201) 933-6300
21
Industrial
Ave
Upper
Saddle
River,
NJ.
07458
P(201)
962
7373
F(201)
8353
E
mail:
[email protected]
Страница 1: ...l No UTG 4000 www phase2plus com 283 Veterans Blvd Carlstadt NJ 07072 201 933 6300 www phase2plus com 21 Industrial Ave Upper Saddle River NJ 07458 P 201 962 7373 F 201 962 8353 E mail info phase2plus...
Страница 2: ...D FUCTIONS 5 3 MEASURING THE THICKNESS 6 3 1 PRESET THE MATERIAL S VELOCITY 6 3 2 PRESET OTHER SPECIFICATIONS 7 3 3 CALIBRATION 8 3 4 DISPLAY MODES 9 3 5 ADJUSTING THE SPECIFICATIONS IN A SCAN MODE 11...
Страница 3: ...wn thickness or sound velocity 1 1 CONSTRUCTION OF THE GAGE 1 2 STANDARD CONFIGURATION Qty Thickness Gauge 1 5MHz PROBE 1 AA Batteries 2 2oz Couplant Gel 1 Rugged Carry case 1 Operation Manual 1 1 3 O...
Страница 4: ...Error 0 001 up to 0 393 steel 0 5 H 0 0003 over 0 393 steel H actual value Display Modes Digital Thickness Readout A scan or B scan Min Max Capture Diff Value Reduction V Path Correction Automatic Up...
Страница 5: ...memory up to 100 000 values 1000 waveforms 10 Selectable Resolution 0 001 0 01 or 01mm 11 Optional rectification mode RF full wave half half 12 Optional waveform outline mode or filled mode 13 Approx...
Страница 6: ...thickness value This part should be measured with a caliper or micrometer 1 Perform a test on the built in test block to verify unit is functioning properly 2 Apply couplant to your sample piece 3 Pre...
Страница 7: ...ctual thickness is less than the minimum value preset MAXIMUM ALARM set the maximum thickness alarm value The result will be displayed in red if the actual thickness is more than the maximum value pre...
Страница 8: ...the method in chapter 3 1 2 After turning on the gage press to enter the calibration interface The screen shows PUT PROBE ON ZERO BLOCK USE COUPLANT 3 Spread the couplant according to the presentatio...
Страница 9: ...erly then the values will be shown in GREEN When the values exceed the upper or lower limit setting then the values will be shown in RED THICKNESS VALUE MODE This interface shows the velocity of mater...
Страница 10: ...VALUE SCANNING MODE this mode allows the customer to test the thickness of material continuously and to show the upper lower limits after the tests It shows the minimum and maximum values during the...
Страница 11: ...are exceeded the letters will be displayed in RED color 3 5 ADJUSTING THE SPECIFICATIONS IN A SCAN INTERFACE In the A scan interface press the P3 button to highlight the value to be adjusted Then pre...
Страница 12: ...ecification is highlighted The first intersection point between the waveform and gate is shown with a RED arrow which will help you determine if this thickness value is correct If correctly tested the...
Страница 13: ...n be realized by adjusting the gate setting above the defect echoes 3 9 REAL CASE 2 3 If there are some surface faults in the testing object and the gate locks the defect echoes the measuring result w...
Страница 14: ...area 12 Gain value 3 72 Introduction to B scan The UTG 4000 has a time base B scan function By moving the probe across the surface of the workpiece you can obtain a cross sectional profile of your par...
Страница 15: ...icates the length of echo blanking the waveform above this is considered invalid See fig 3 16 below 1 Identification of dual echo mode 2 Blue arrow indicates secondary echo 3 E blanking 4 Blue line th...
Страница 16: ...to main menu 6 Save current value or waveform 7 Delete the selected data 4 2 Browsing stored data Viewing Individual Files Press M to enter the configuration display Highlight GRID FILE and press the...
Страница 17: ...ful attention to polarity Attention when not using the gauge for extended periods of time please remove batteries to prevent any leakage or corrosion 5 2Maintenance 1 Always keep your gauge probe cabl...
Страница 18: ...on a sample piece of known thickness which is at or near the temperature of the material to be measured This will allow the gauge to correctly calculate the velocity of sound through the hot material...
Страница 19: ...osity liquid is used as the medium The sound used in ultrasonic thickness measurement does not travel through air efficiently A wide variety of couplant materials may be used in ultrasonic gauging Pro...
Страница 20: ...te 0 1060 2680 Molybdenum 0 2460 6250 Motor oil 0 0690 1740 Nickel pure 0 2220 5630 Polyamide 0 0870 2200 Nylon 0 1020 2600 Polyethylene high density HDPE 0 0970 2460 Polyethylene low density LDPE 0 0...
Страница 21: ...t Beijing 100192 China Tel 86 10 59792409 Fax 86 10 59814851 General E mail info phase2china com cn MEXICO Phase II de Mexico Calle A No 4 Promer Piso Col San Marcos Azcapotzalco C P 02020 Mexico Tel...
Страница 22: ......