 Manual Tablet Testing Instrument 3-in-1 - Type PTB 311E 

 

 

  29-02860_09.1.0E 

Page 1 of 5 

PHARMA TEST AG 

 

  

+49 6182 9532-600

Siemensstrasse 5

 

 

 

 

   

+49 6182 9532-650

 

D-63512 Hainburg (GER)

   

 

 

 [email protected] 

 

 

 

 

 

 

 www.pharma-test.com

 

 

 
The Manual Tablet Testing Instrument 

PTB 311E

 is a 

dual

 force mode hardness test 

apparatus as it can be used for either linear force or linear speed increase while Tablet 
Hardness is tested. It offers a multiple point validation procedure for the built-in digital load 
cell and the diameter, height station. 
 
Test Thickness, Diameter or Length and Hardness of the same Tablet using just one 
Instrument, the PTB 311E. The instrument is made in strict compliance with the EP <2.9.8> 
and USP <1217> Pharmacopoeia. 
 
Enter the nominal test information for Thickness, Diameter or Length and Hardness via the 
PTB 311E keyboard. Select the unit to measure, mm - IN - kp, N or Sc, now place the 
sample upright into the Thickness Test Holder and start the test. The driven jaw will now run 
forward and touch the tablets highest point to measure the thickness, thereafter it reverses 
back so that the sample falls into horizontal position to test diameter and hardness. 
 
The 3 results are immediately displayed and printed to a connected Dot-Matrix or suitable 
PCL Type Laser- or DeskJet printer. Repeat this until your series has been tested, get a full 
print including each individual result, mean value and deviations. 
 
In case you need to test hardness “only”, no problem; zero Thickness and Diameter station 
and do hardness test only. This flexibility and the reproducibility of the results have made the 
PTB311E series to become one of the most sold hardness testers. 

 
Operating Principle… 

 
Even in the existing monographs from USP and EP is no standard force setting or force 
increase mode established. This usually is causing problems comparing results received of 
different suppliers instruments of the same tablet. The hardness result is directly influenced 