background image

Operation

17

6 Operation

6.1 Before switching on the pump

Check the operating fluid level in the sight glass.

Compare the voltage and frequency information on the rating plate with the mains volt-
age and frequency values.

Check that the exhaust connection allows free flow (max. permissible pressure 150  
kPa absolute).
– Activate the shut-off valves in such a way that they open before or at the same time 

as the pump is started.

Protect the pump sufficiently from taking in contaminants by means of suitable precau-
tions (e.g. dust filters); if necessary, check operating fluid regularly or replace at short-
er intervals.

6.2 Switching on the pump

The pump can be switched on in any pressure range between atmospheric and
ultimate pressure.
No special precautions are necessary when pumping dry gases. In order to attain the 
lowest possible ultimate pressures, the gas ballast valve should be closed.

Switch on pump at main switch 15.

Switch on the pump with the vacuum flange closed and allow to warm up for 30 min-
utes.

6.3 Pumping condensable vapours

Should the process gases contain condensable gases present at high percentages, the 
rotary vane pump must be operated with gas ballast (i.e. with an open gas ballast valve).

Gas ballast valve, 
standard version

To avoid condensation in the pump when pumping condensable vapours, air is periodi-
cally fed into the working chamber at the beginning of the compression phase via the gas 
ballast valve 42.
The gas ballast valve is closed when turning to the right to position 0 and open when turn-
ing to the left to position 1. Intermediate settings are not possible.

CAUTION

Hot surface!

Danger of burns if hot parts are touched. Depending on the operating and ambient con-
ditions, the surface temperature of the pump may rise above 70 °C.

In this case, use suitable finger guards.

NOTICE

Bad final vacuum and damage to the pump!

Danger of condensation and corrosion due to exceeding the water vapour compatibility 
during operation without gas ballast or in case of insufficient supply of flushing gas.

Only pump vapours when the pump is warm and the gas ballast valve is open.

When the process has been completed, allow the pump to continue running for 
about 30 minutes with the vacuum flange closed and the gas ballast open for ope-
rating fluid regeneration purposes.

Содержание DUO 20 M

Страница 1: ...Translation of the original instructions Operating Instructions EN PK 0203 BEN L 1208 DUO 20 M DUO 20 MC Rotary Vane Pump...

Страница 2: ...he mains power supply 11 5 5 Filling up the operating fluid 15 5 6 Operations monitoring Option 16 6 Operation 17 6 1 Before switching on the pump 17 6 2 Switching on the pump 17 6 3 Pumping condensab...

Страница 3: ...ng instructions are the result of risk evaluations and hazard analyses and are oriented on international certification stan dards as specified by UL CSA ANSI Z 535 SEMI S1 ISO 3864 and DIN 4844 In thi...

Страница 4: ...Exhaust flange Gas ballast valve Power connection Prohibition of an action or activity in connection with a source of danger the disregarding of which may result in serious accidents Warning of a disp...

Страница 5: ...mbled components of the magnetic coupling separate at all times Danger of crushing Do not allow any magnetised parts into the vicinity of the magnetic coupling Danger of injury 2 2 Protective equipmen...

Страница 6: ...atibility levels of the pump pumping of substances that tend to sublime use of the vacuum pump to generate pressure pumping of liquids the use of operating fluids not specified by Pfeiffer Vacuum conn...

Страница 7: ...e top side of the pump to lift the pump Is the pumps weight 25 kg lift the pump by a crane Fig 1 Transporting the pump 3 2 Storage Check that all the openings on the pump are securely closed Store the...

Страница 8: ...y Pump with motor power cord power switch dependent on the motor type Operating fluid P3 for standard pump Cone strainer and centering ring with O ring Locking cap for vacuum and exhaust flange Operat...

Страница 9: ...hamber vacuum tight and at the same time vents the pump The integrated magnetic coupling functions free of contact and friction Therefore it is not subjected to any mechanical wear and is completely m...

Страница 10: ...t be visible 5 2 Connecting the vacuum side Remove locking cap from the vacuum flange pay attention to the cone strainer and the respective O ring in the intake port The connection between the pump an...

Страница 11: ...poisoning from emitted gases or vapours which can be detrimental to health and or can pollute the environment depending on the particular application Comply with the applicable regulations when worki...

Страница 12: ...with slow characteristics The transmission power of the pump s magnetic coupling is so great that there is no overload protection for the motor NOTICE Overvoltage An incorrect voltage range setting c...

Страница 13: ...is the cube root of the coil voltage the mains and the coil current are the same Star connec tions are denoted by the symbol Y Fig 7 Motor coil and connecting plate of Star Connection for high voltag...

Страница 14: ...tions so that the directional rotation indicated on the pump is main tained regardless of the representations in the current flow diagram Fig 8 Connection example for a three phase AC motor with PTC r...

Страница 15: ...ing operations within the markings at the sight glass frame Fig 9 Filling up the operating fluid Screw in operating fluid filler screw 198 Check operating fluid level only when the pump is warm and ru...

Страница 16: ...h opens The signal can be used to con trol external valves Fig 10 Installation location and circuit diagram of pressure switch Switch off the pump Unscrew and remove locking screw 142 on the support R...

Страница 17: ...ld the process gases contain condensable gases present at high percentages the rotary vane pump must be operated with gas ballast i e with an open gas ballast valve Gas ballast valve standard version...

Страница 18: ...s at the flushing gas connection Set flushing gas pressure maximum pressure 1 5 bar absolute Select the type and amount of flushing gas depending on the process consult Pfei ffer Vacuum if necessary 0...

Страница 19: ...Maintaining the vac uum in the chamber NOTICE Danger of backflow of operating fluid into the intake line Contamination of the connected vacuum system Vent the vacuum chamber within 30 s regardless of...

Страница 20: ...is disassem bled Persons with cardiac pacemakers must not enter the area 2m of the magnetic field Rooms in which open couplings are accessible must be identified No trespassing for persons with heart...

Страница 21: ...rs as required at least every 4 years Check operating fluid level X Visual inspection leak tightness oil leaks X Check filter insert of external oil mist filter if existent X Change filter insert of e...

Страница 22: ...a Fig 13 Draining the operating fluid Screw in operating fluid drain screw 198a pay attention to O ring Screw in operating fluid filler screw 198 Depending on the applications Pfeiffer Vacuum recommen...

Страница 23: ...of the sight glass Operate the pump with the gas ballast open until the pump has warmed up Drain the operating fluid again and check for contamination flush again if necessary Screw the operating flui...

Страница 24: ...mantling Unscrew 2 cheesehead screws 182 from the gas ballast valve housing Remove the gas ballast valve housing 42 from the casing taking care with O ring 148 Unscrew silencer nozzle 44 and O ring 13...

Страница 25: ...ing the pump Do not store pump in the vicinity of machines lanes etc because strong vibrations can damage the rotor bearings 8 2 Re starting 8 3 Disposal Products or parts thereof mechanical and elect...

Страница 26: ...ure of the pump may rise above 105 C in case of malfunction Carry out work on the pump only after it has cooled to a safe temperature NOTICE Motor overload Depending on the malfunction e g blocking du...

Страница 27: ...mponents for con tamination Operating fluid dirty Operate pump for a longer period with gas ballast valve open or change operating fluid Leak in system Repair leak Operating fluid filling level too lo...

Страница 28: ...t in the shipment required Dismantle all accessories Drain operating fluid lubricant Drain cooling medium if used Send the pump or unit in its original packaging if possible Sending of contaminated pu...

Страница 29: ...from A PK E32 011 T PK E30 008 T 7 8 9 10 32 44 51 102 106 110 114 193 195 DUO 20 MC without as from A PK E32 013 T PK E30 008 T 7 8 9 10 32 44 51 102 106 110 114 193 195 Vacuum safety valve DUO 20M D...

Страница 30: ...ate 104 Valve trap 106 Hydraulic piston 108 Compression spring 110 Sight glass 112 Sight glass seal 114 Rubber foot 116 Flange 138 O ring 141 O ring 144 O ring 147 O ring 148 O ring 149 O ring 152 O r...

Страница 31: ...ing oil pressure switch PK 196 484 T Barretter actuation unit 3TF P 4768 051 FQ Operations monitoring unit 3 operating fluid level operating fluid temperature ex haust pressure PK 196 141 T Operations...

Страница 32: ...Yes with 2 m cable and safe ty plug Yes with 2 m mains cable and safety plug Operating fluid filling 1 1 l 1 1 l Weight 43 5 kg 43 5 kg Parameter DUO 20 M 3 phase motor 3TF DUO 20 MC 3 phase mo tor 3T...

Страница 33: ...112 140 7 7 3 178 113 238 A 252 H 18 1 13 134 166 8 99 77 5 266 113 191 DN 25 ISO KF 5 76 5 Pump with three phase motor Pump with single phase motor A 542 mm 547 5 1 mm H 212 without motor protection...

Страница 34: ...ble for compiling the technical documentation is Mr Sebastian Ober beck Pfeiffer Vacuum GmbH Berliner Stra e 43 35614 A lar DuoLine DuoLine DUO 20 M MC Guidelines harmonised standards and national sta...

Страница 35: ...ngineering art competent advice and reliable services Ever since the invention of the turbopump we ve been setting standards in our industry And this claim to leadership will continue to drive us in t...

Отзывы: