14
Connection "DeviceNet"
Baud rate
The DeviceNet baud rate is set manually using the "RATE" selector switches or via De-
viceNet.
Setting the baud rate manually
Set the selector switches to the desired value.
– The change takes effect the next time the system is switched on.
Setting the baud rate via DeviceNet
Switch off the pump/electronic drive unit
Set the selector switches to "P".
– When switched on, the system uses the last valid baud rate (500 kBit/s upon deliv-
ery).
– The baud rate is programmable via DeviceNet object 3.1.2.
6.3 Configuring the data exchange
Depending on the program used to set up DeviceNet communication, you can carry out
the following steps:
Import EDS file (see scope of delivery)
Device-specific data:
Specify the format of the cyclical input/output data
– Specify the cyclically alternating process data using objects "Poll I/O input data
(4.0.100)" and "Poll I/O output data (4.0.101)"
– The format can be changed only if the poll I/O connection is inactive.
Specify system behaviour following termination of the cyclical DeviceNet communica-
tion.
– Use object "Idle Action (9.1.7)" to specify which action is to be carried out if there
is a failure of the cyclical exchange of process data (poll I/O connection).
Specify the remote priority.
– Use object "Permission locked (101.0.17)" to specify whether the electronic drive
unit is to be controlled exclusively via DeviceNet or whether other interfaces (e.g.
RS485) are also permitted.
– This does not restrict write access via other interfaces.
Perform pump configuration.
– To set a configuration that differs from the delivery state, use the EDS file to adjust
individual objects (e.g. accessory configuration).
Master-specific data:
Set up the device in the master scanlist.
Set the format of the cyclical input/output data
Explicit data ex-
change (explicit con-
nection)
This connection can be used to access the individual DeviceNet objects
. As a rule, this is done using a dedicated configuration program and the EDS file.
This is also used to specify which data is sent during the cyclical data exchange.
RATE
P
1
2
5
Position
Meaning
1
125 kBit/s
2
250 kBit/s
5
500 kBit/s
P
Baud rate via DeviceNet