OPERATING INSTRUCTIONS
EN
Translation of the Original
TC 110
Electronic drive unit
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original TC 110 Electronic drive unit...
Страница 2: ...fer vacuum de Further operating instructions from Pfeiffer Vacuum can be found in the Download Center on our website Disclaimer of liability These operating instructions describe all models and varian...
Страница 3: ...4 4 1 Connection diagram 14 4 2 Connection X3 16 4 2 1 Voltage supply 17 4 2 2 Inputs 17 4 2 3 Outputs 18 4 2 4 RS 485 18 5 Interfaces 19 5 1 Interface RS 485 19 5 1 1 Connection options 19 5 1 2 Cros...
Страница 4: ...set value 36 7 2 8 Backing pump operating modes 36 7 2 9 Backing pump standby mode 37 7 2 10Operation with accessories 37 7 2 11Venting modes 38 7 3 Switching on the turbopump 38 7 4 Switching off th...
Страница 5: ...ation and meaning of the parameters 23 Tbl 11 Control commands 26 Tbl 12 Status requests 27 Tbl 13 Set value settings 28 Tbl 14 Parameter for DCU functions 29 Tbl 15 Configure parameters P 062 and P 0...
Страница 6: ...110 with external wiring 16 Fig 5 Connection of the electronic drive unit via X3 19 Fig 6 Cross link via connection X3 using connection cables 20 Fig 7 Schematic diagram of power characteristics examp...
Страница 7: ...ons performing the following activities on the product transport setup installation usage and operation decommissioning maintenance and cleaning storage or disposal The work described in this document...
Страница 8: ...DCU Display Control Unit from Pfeiffer Vacuum DI DO Digital input digital output f Rotation speed value of a vacuum pump frequency in rpm or Hz HPU Handheld Programming Unit I Electric current LED Li...
Страница 9: ...iding property damage Notes tips or examples indicate important information on the product or on this document 2 2 Safety instructions All safety instructions in this document are based on the results...
Страница 10: ...n on potential dangers The product holder or user is obliged to make all operating personnel aware of dangers posed by this product Every person who is involved in the installation operation or mainte...
Страница 11: ...clusively for the operation of Pfeiffer Vacuum turbopumps and their accessories 2 6 Foreseeable improper use Improper use of the product invalidates all warranty and liability claims Any use that is c...
Страница 12: ...use in safety related applications in accordance with EN ISO 13849 1 No proof test is required throughout the expected device lifetime of up to 20 years If you calculate your safety application with...
Страница 13: ...4 V DC Connection panel Standard X3 Turbopump HiPace 10 30 60 80 300 Splitflow 50 80 Tbl 6 Features of the device variants 3 3 Function 1 2 3 Fig 1 TC 110 connection panel 1 PV can service connection...
Страница 14: ...When establishing the voltages that exceed the specified safety extra low voltage according to IEC 60449 and VDE 0100 the insulating measures will be destroyed There is a danger to life from elec tric...
Страница 15: ...S2 S1 Fig 2 Connection diagram for TC 110 with connection cable A B C 3 4 2 1 UB PE A B C DC in 1 2 3 4 5 24 VDC RS 485 RS 485 n c RS485 2 3 4 5 6 7 10 11 8 9 12 13 14 15 1 1 8 9 15 X3 DI DI1 DI2 DI D...
Страница 16: ...he X3 plug and socket connection with voltage supply switched on may lead to the de struction of electronic components Always interrupt the voltage supply before disconnecting the X3 connecting plug S...
Страница 17: ...s Tbl 8 Terminal layout of the 15 pin connection X3 4 2 1 Voltage supply 24 V DC input pin 1 The electrical connection to X3 is made using connection cable from the Pfeiffer Vacuum accessory program o...
Страница 18: ...igh After reaching the rotation speed switch point Rotation speed switch point 1 has a factory setting of 80 of the nominal rotation speed This can be used for example for a Turbopump ready for operat...
Страница 19: ...unit DCU or HPU or an external computer The connections are galvanically safe and are isolated from the maximum supply voltage for the electronic drive unit The electrical connections are optically de...
Страница 20: ...verter Networking as RS 485 Bus The group address of the electronic drive unit is 961 1 Install the devices according to the specification for RS 485 interfaces 2 Make sure that all devices connected...
Страница 21: ...n0 l1 l0 dn d0 c2 c1 c0 cR Error message a2 a1 a0 1 0 n2 n1 n0 0 6 N O _ D E F c2 c1 c0 cR _ R A N G E _ L O G I C NO_DEF Parameter number n2 n0 no longer exists _RANGE Data dn d0 outside the permiss...
Страница 22: ...lse true 06 000000 corresponds with false 111111 corresponds with true 1 u_integer pos whole number 06 000000 to 999999 2 u_real pos fixed point number 06 001571 corresponds with 15 71 4 string Charac...
Страница 23: ...ons Function description of the parameters Data type Type of formatting of the parameter for the use with the Pfeiffer Vacuum protocol Access type R read Read access W write Write access Unit Physical...
Страница 24: ...heating 13 backing pump 14 sealing gas 15 pumping station 16 pump rotates 17 pump does not rotate 19 pressure switch point 1 underrun 20 pressure switch point 2 underrun 21 fore vacuum valve is de cel...
Страница 25: ...controlled 7 RW 0 14 0 036 Cfg Acc B1 Configura tion accesso ry connec tion B1 Functions see P 035 7 RW 0 14 1 037 Cfg Acc A2 Configura tion accesso ry connec tion A2 Functions see P 035 7 RW 0 14 3 0...
Страница 26: ...RW 0 7 2 Tbl 11 Control commands 6 3 Status requests Display Description Func tions Data type Access type Unit min max de fault 300 RemotePrio Remote priority 0 no 1 yes 0 R 0 1 302 SpdSwPtAtt Rotati...
Страница 27: ...ic drive unit 4 R 354 HW Version Hardware version electron ic drive unit 4 R 360 ErrHist1 Error code history item 1 4 R 361 ErrHist2 Error code history item 2 4 R 362 ErrHist3 Error code history item...
Страница 28: ...R 740 Pressure 1 Pressure value 1 10 RW hPa 742 PrsCorrPi 1 Correction factor 1 2 RW 749 PrsSn2Name Name sensor 2 4 R 750 Pressure 2 Pressure value 2 10 RW hPa 752 PrsCorrPi 2 Correction factor 2 2 R...
Страница 29: ...3 350 Ctr Name Display and control panel type 4 R 351 Ctr Software Display and control panel software version 4 R 738 Gauge type Type of pressure gauge 4 RW 794 Param set Parameter set 0 Basic pa rame...
Страница 30: ...ached 1 No error active with trouble free operation 2 Error active if the error message is active 3 Warning active if a warning message is active 4 Error and or warning active if an error and or warni...
Страница 31: ...Fore vacuum control Fore vacuum side Control of Pfeiffer Vacuum pumping stations Tbl 17 Configure parameter P 055 7 1 4 Configuring the accessory connections Connect accessory devices to the TC 110 T...
Страница 32: ...erface serves the display of the interface currently active in the electronic drive unit The communication interfaces thus automatically achieve control priority Option Description 1 remote Operation...
Страница 33: ...maximum power consumption When the nominal respective set rotation speed is reached the electronic drive unit automatically switches over to the chosen power characteris tic of the selected gas mode...
Страница 34: ...302 Rotation speed switch point 1 Process f t t t P 010 P 302 P 701 P 017 0 0 1 0 1 Fig 8 Rotation speed switch point 1 active Adjusting rotation speed switch point 1 Signal output and status paramet...
Страница 35: ...ation speed Values in the rotation speed setting mode or standby mode are subject to the permissible rotation speed range of the respective vacuum pump technical data Underrunning the minimum permissi...
Страница 36: ...cking pump type Operating mode P 025 Recommended booster pump 0 Continuous operation All backing pumps 1 Interval mode Diaphragm pumps only 2 Delayed switching on All backing pumps 3 Delayed interval...
Страница 37: ...ng a switching delay depending on the process or application requirements The delay is depending on a permanently uninterrupted exceedence or undershot of the specified switching threshold Switch off...
Страница 38: ...e venting mode of the turbopump after switching off The sig nal output is carried out with a fixed delay Selecting the venting mode 1 Set parameter P 012 to 1 2 Select the venting mode with parameter...
Страница 39: ...unit show the basic operating states of the vacuum pump A differentiated error and warning display is only possible for operation with the Pfeiffer Vacuum display and control unit or a PC LED Symbol...
Страница 40: ...temperature Examples are an impermissible motor temperature or impermissibly high housing tempera ture Further reduction of drive power and thus decreasing speed can potentially lead to underrun the...
Страница 41: ...ched off Warnings Warning F do not cause components to switch off Handling malfunction messages 1 Read out the error codes via the display and control panel or the PC 2 Remove the cause of the malfunc...
Страница 42: ...rvice Only acknowledge when rotation speed f 0 Err112 Operating fluid pump group error Contact Pfeiffer Vacuum Service Only acknowledge when rotation speed f 0 Err114 Final stage temperature evaluatio...
Страница 43: ...m Service Wrn100 Rotation speed raised to minimum value Permissible specifications for ro tation speed operation or stand by incorrect Check P 707 or P 717 Obtain the valid rotation speed range from t...
Страница 44: ...acuum Service Error E042 Hardware error EPROM checksum incorrect Contact Pfeiffer Vacuum Service Error E043 Hardware error E2 PROM write error Contact Pfeiffer Vacuum Service Error E090 Internal devic...
Страница 45: ...and addresses can be found on our website in the Pfeiff er Vacuum Service section Advice on the optimum solution is available from your Pfeiffer Vacuum contact partner For quick and smooth handling of...
Страница 46: ...ontainers e Observe the applicable transport conditions ERKL RUNG KONTAMINIERUNG 6 Affix the declaration of contamination to the outside of the packag ing 7 Then send your product to your local Servic...
Страница 47: ...Restriction of the use of certain hazardous substances 2011 65 EU TC 110 Harmonized standards and applied national standards and specifications DIN EN 61000 3 2 2014 DIN EN 61000 3 3 2013 DIN EN 6101...
Страница 48: ...PT0204 U Date 1903 P N PT0204BEN...