Installation
9
5.2
Installing the operating fluid return line
Unscrew operating fluid drain screw 33.
Drain off operating fluid.
Separate the operating fluid from the condensate.
– If the drained operating fluid is free of contamination, it can be reused.
UnoLine, DuoLine
OME 16 M
Fig. 2:
Duo 3 with OME 16 M and ODK
Screw in hose connection 221 in place of the operating fluid drain screw 33; take care
with O-ring 145.
Unscrew and remove the union nut of the screw fitting 28 and remove the sealing plug
29.
Fit operating fluid return hose 222 at both sides,
– keep the hose as short as possible and protect it from bending.
Tighten the union nut of the screw fitting 240.
Fasten hose clip 223.
NOTICE
Make sure that the return for the operating fluid is working properly.
The operating fluid can only be sucked out and returned when there is a minimum about
of operating fluid in the oil mist eliminator.
If necessary, top up with operating fluid in order to ensure the return of operating fluid
at the start of the evacuation phase.
Pour the operating fluid slowly into the outlet flange of the oil mist eliminator until it
can be seen in the sight-glass.
33
Drain screw for operating fluid
28
Screw fitting
29
Closure cap
145 O-ring
221 Hose
222 Operating fluid return
hose
223 Hose clamp