Display
Designations
Functions
Data
type
Ac-
cess
type
Unit min. max. de-
fault
020 PressMode
Pressure regu-
lation
0 = off
1 = on
0
RW
0
1
0
024 Cfg DO1
Output DO1
configuration
0 = Switch-point reached
1 = No error
2 = Error
5 = Set rotation speed
reached
6 = Pump on
9 = Always "0"
10 = Always "1"
11 = Remote priority active
21 = Gas ballast
7
RW
0
21
21
026 SpdSet-
Mode
Rotation speed
setting mode
0 = off
1 = on
7
RW
0
1
0
030 VentMode
Valve mode
0 = auto (factory setting)
1 = closed
2 = open
7
RW
0
2
0
034 Auto start
automatic start
after power
failure
0 = off
1 = on
0
RW
0
1
0
052 BalGasValv
Gas ballast
valve control
0 = off
1 = on
0
RW
0
1
0
060 CtrlViaInt
Operate via in-
terface
1 = remote
2 = RS-485
255 = Unlock interface selec-
tion
7
RW
1
255
1
061 IntSelLckd
Interface se-
lection locked
0 = off
1 = on
0
RW
0
1
0
063 Cfg DI2
Configuration
input DI2
0 - 4 = Deactivated
5 = Rotation speed setting
mode On
7
RW
0
5
0
068 Cfg Acc C1
Configuration
accessory con-
nection C1
2)
15 = gas ballast valve
16 = vacuum safety valve ,
type IV 16 M
17 = pressure sensor
3)
18 = vacuum safety valve,
type ISV 25
7
RW
0
7
0
069 Cfg Acc D1
Configuration
accessory con-
nection D1
4)
15 = gas ballast valve
16 = vacuum safety valve ,
type IV 16 M
17 = pressure sensor
18 = vacuum safety valve,
type ISV 25
7
RW
0
7
0
095 FactorySet
Factory set-
tings
1 = Reset to factory configura-
tion
0
RW
0
1
0
Tbl. 10:
Control commands
2)
Detection through interface
3)
Preferred plug-in position
4)
Detection through interface
Parameter set
39/84