15.1.2 Quadrupole controller QC 700
Parameter
Value
Suitable for
System chassis SC 700
Required plug-in positions
2
Quantity per system
1
Plug-in positions
Plug-in positions no. 1 and 2 (on the very left in system chassis SC 700)
Weight
0.4 kg
Tbl. 21:
Technical data, QC 700
Parameter
Value
Number of measurement channels
128
Operating modes
MONO / MULTI channel
Measurement cycles
1 – 10,000 or REPEAT
Channel change time
100 – 200 μs (with min. PAUSE in the cycle)
Tbl. 22:
Operating modes and parameters QC 700
Mass MODE
Measuring method
SCAN-N
Analog scan, normal
SCAN-F
Analog scan with FIR filter for the measured value
STAIR
Scan bargraph
SAMPLE
Single ground and MID
PEAK-L
Peak search with level criterion
PEAK-F
Peak search with FIR filter
ADJUST-C
Rough search
ADJUST-F
Fine search
Tbl. 23:
Mass flow rate operating modes QC 700
Scan SPEED
STEPS per ground
FIX Range
AUTO Range
125 μs/u
8/u
---
250 μs/u
16/u
---
0.5 – 1 ms/u
32/u
---
2 – 20 ms/u
64/u
32/u
50 ms/u – 60 s/u
64/u
64/u
Tbl. 24:
Mass scale resolution QC 700
Mass MODE
EP 422 or external inputs
Ion counter (CP 400)
FIX Range
AUTO Range
AUTO Range
SAMPLE
125 μs – 60 s
0.5 ms – 60 s
125 μs – 60 s
STAIR
125 μs/u – 60 s/u
2 ms/u – 60 s/u
0.5 ms/u – 60 s/u
SCAN
125 μs/u – 60 s/u
5 ms/u – 60 s/u
1.25 ms/u – 60 s/u
PEAK
125 μs/u – 60 s/u
5 ms/u – 60 s/u
1.25 ms/u – 60 s/u
Tbl. 25:
Measurement speeds QC 700
Technical data and dimensions
62/70