Accessories

11

Continous operation

With "pumping station on", the electronic drive unit sends a signal to the configured ac-
cessory connection to switch on the backing pump. This signal can also be used for con-
trolling a fore-vacuum safety valve.

Intermittend opera-
tion (diaphragm 
pumps only)

Intermittend operation can extend the life expectancy of the membrane of a connected 
diaphragm pump. Either a diaphragm pump with built-in semiconductor relay or an inter-
connected relay box with semiconductor relay is required for intermittend operation. The 
backing pump is switched on and off in dependence of the turbopump's power consump-
tion. A relation to the supplied fore-vacuum pressure is derived from the power consump-
tion. The switching off and switching on thresholds for the backing pump are adjustable. 
Fluctuations in the power consumption of idling turbopumps and type-dependent varying 
fore-vacuum pressures of the backing pumps require the switching thresholds to be set 
separately for the intermittend mode.
Pfeiffer Vacuum recommends the intermittend mode between 5 and 10 hPa. A pressure 
gauge and a dosing valve are required to set the switching thresholds.

Switch on the vacuum system via the function "pumping station" and await the run-up.

Generate a fore-vacuum pressure of 10 hPa by gas inlet via dosing valve.

Read and note the parameter 

[P:316]

Adjust the switch on threshold backing pump via parameter 

[P:711]

 to the determined 

drive power for a fore-vacuum pressure of 10 hPa.

Reduce the fore-vacuum pressure to 5 hPa.

Read and note the parameter 

[P:316]

Adjust the switch off threshold backing pump via parameter 

[P:710]

 to the determined 

drive power for a fore-vacuum pressure of 5 hPa.

Delayed switching on

Switching on the turbopump and the backing pump at the same time can result in un-
wanted gas flows. Depending on process or application requirements, the backing pump 
can be switched on with a delay. The switch-on delay depends on the rotation speed of 
the turbopump and is fixed in the electronic drive unit at 6 Hz.
The switch-on delay signal can also be used for switching a fore-vacuum safety valve.

7

Accessories

An overview about original Pfeiffer Vacuum accessories for the designated device can 
be found in the operating instructions of the respective vacuum pump.

8

Technical data

Parameter

Relay box, shielded, 
for backing pumps

Relay box, shielded, 
for backing pumps

Mains requirement: frequency (range)

50/60 Hz

50/60 Hz

Mains requirement: voltage (range)

100-240 (± 10 %) V AC

100-240 (± 10 %) V AC

Contact rating

7 A

20 A

Control voltage

24 V DC

24 V DC

Содержание HiPace PM 071 282-X

Страница 1: ...Translation of the original instructions Operating Instructions EN PT 0444 BEN A 1308 HiPace Backing pump relay box shielded...

Страница 2: ...e 5 3 Product description 6 3 1 Relay box single phase 7 A 6 3 2 Relay box single phase 20 A 6 4 Electrical connection 7 5 Connection to the electronic drive unit 9 5 1 TC 110 9 5 2 TC 400 TM 700 9 5...

Страница 3: ...Z 535 SEMI S1 ISO 3864 and DIN 4844 In this document the following hazard levels and information are considered Backing pump relay boxes Order number Relay box shielded for backing pumps 1 phase 7A fo...

Страница 4: ...onditions which afford other protection classes Always ensure a safe connection to the protective earthing conductor PE protection class I Before carrying out any work disconnect the unit and all asso...

Страница 5: ...1200 with Y Connector external electronic drive unit TCP 350 with adapter cable The shielded version supports the electromagnetic compatibility and reduces interac tion with other electric devices 2...

Страница 6: ...um diaphragm vacuum pumps The maximum relay contact load is 7 A 3 2 Relay box single phase 20 A The relay box PM 071 283 X PM 071 285 X with mechanical relay is provided for the operation of larger Pf...

Страница 7: ...r installation Electrical connection may be carried out only by trained and authorised electricians Ensure the system is adequately earthed Establish an adequate fuse protection on customer side depen...

Страница 8: ...fittings Close the relay box 1 1a 12 1 1 11 11 S h i e l d AC in AC out 20 L1 N PE PE N L 3 1c 1b Fig 4 Relay box for backing pump single phase 20 A 1 Casing 1 1 Casing cover 1a Control lead for elect...

Страница 9: ...T or via TCS 12 Make the settings and control via the interfaces of the electronic drive unit 5 2 TC 400 TM 700 Connect the control lead of the relay box directly to a vacant accessory port of the el...

Страница 10: ...s of the electronic drive unit 6 Operation Establish the mains supply for the relay box Make sure the supply voltage for the backing pump is valid Observe the rating plates 6 1 Operation mode backing...

Страница 11: ...he vacuum system via the function pumping station and await the run up Generate a fore vacuum pressure of 10 hPa by gas inlet via dosing valve Read and note the parameter P 316 Adjust the switch on th...

Страница 12: ...12 Technical data 8 1 Dimensions Fig 5 Dimensions of the relay boxes 7 A and pattern of fixing holes 125 113 52 80 57 42 10 4 8 9 8...

Страница 13: ...Technical data 13 Fig 6 Dimensions of the relay boxes 20 A and pattern of fixing holes 120 122 91 61 20 7 11 10 5 82 106...

Страница 14: ...ty 2004 108 EC PM 071 282 X PM 071 283 X PM 071 284 X PM 071 285 X HiPace Harmonised standards and national standards and specifications which have been ap plied DIN EN 61000 3 2 2008 DIN EN 61000 3 3...

Страница 15: ...e the only supplier of vacuum technology that provides a complete product portfolio Benefit from our know how and our portfolio of training opportunities We can support you with your plant layout and...

Отзывы: