10
Operation
5.3
TC 1200
Connect the control lead of the relay box directly to a vacant accessory port of the
electronic drive unit
or
using an Y-Connector.
Make the settings and control via the interfaces of the electronic drive unit.
5.4
TCP 350
Connect the control lead of the relay box to the "remote" connection of the electronic
drive unit using an adapter cable.
Make the settings and control via the interfaces of the electronic drive unit.
6
Operation
Establish the mains supply for the relay box.
– Make sure the supply voltage for the backing pump is valid.
– Observe the rating plates!
6.1
Operation mode backing pump
Operation of a connected backing pump via the electronic drive unit depends on the
backing pump type.
Adjust the parameter
[P:025]
to the desired value.
V
1a
TC 1200
1b
1c
1
V
1c
PM 061 376 -T
PM 061 377 -T
TCP 350
1b
1a
1
Operation mode [P:025]
recommended backing pump
"0" continous operation
all kinds of backing pumps
"1" Intermittend operation
diaphragm pumps only
"2" Delayed switching on
all kinds of backing pumps