Removing the operating fluid reservoir
1. Wear laboratory gloves to avoid skin contact.
2. Place the turbopump on the closed high vacuum flange.
3. Unscrew all Allen head screws from the screw cap on the pump base.
4. Remove the screw cap.
5. Remove the O-ring from the groove using an O-ring picker.
– Avoid damage caused by scratches.
6. Remove the operating fluid reservoir from the bearing housing by hand.
7. Pull the old capillary rods out of the pump base with the tweezers.
8. Clean the screw cap with a clean, lint-free cloth.
– Do
not use cleaning agents
.
7.3.2 Installing the operating fluid reservoir
Required consumables
● Laboratory gloves
● Operating fluid reservoir
Required tools
● Allen key,
WAF 5
● Tweezers
● Calibrated torque wrench (tightening factor ≤ 1.6)
4
5
6
1
2
3
7
Fig. 21:
Installing the operating fluid reservoir
1 Operating fluid reservoir
5
Screw cap
2 Capillary rods (9×)
6
O-ring
3 Injection tip
7
Protective cover
4 Allen head screw
Installing the operating fluid reservoir
1. Wear laboratory gloves to avoid skin contact.
2. Insert all new capillary rods with the tweezers.
3. Insert the operating fluid reservoir into the bearing housing with the felt side in the direction of the
injection tip.
– When doing so, do
not
exert pressure on the operating fluid reservoir.
4. Insert the O-ring in the groove in the pump base.
5. Fit the screw cap.
6. Tighten all 3 Allen head screws uniformly.
– Tightening torque:
2.5 Nm
.
Maintenance
43/62
Содержание HiPace 300 P
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original HIPACE 300 P Turbopump ...
Страница 54: ...13 Accessories View the range of accessories for hybrid bearing turbopumps on our website Accessories 54 62 ...
Страница 61: ...61 62 ...
Страница 62: ...Notizen Notes ...
Страница 63: ......
Страница 64: ... PT0633 ed C Date 2108 P N PT0633BEN ...