Procedure
1. Remove the protective cap from the exhaust connection.
2. With vertical installation, connect the exhaust line directly with the exhaust connection.
3. With horizontal installation (180°), connect the exhaust line using an outlet elbow.
4. Route the piping downwards from the vacuum pump, to prevent condensate return.
5. Install a condensate separator if necessary.
5.4 Connecting cooling water supply
Optional monitoring systems
● Flow rate monitor
● Cooling water pressure monitor
Prerequisites
● Cooling water connection unpressurised
● Cooling system evacuated
1
2
Fig. 6:
Cooling water connection
1 Cooling water inlet
2
Cooling water outlet
Connecting cooling water supply
1. Connect the cooling water connections with the water supply.
– Connection size: 1/2" thread, ISO 228-1
2. Open the water inflow.
Parameter
Cooling water
Appearance
● filtered
● mechanically clear
● visually clear
● no turbidity
● no sediment
● free from grease and oil
pH value
7 to 8
Carbonate hardness, max.
10 °dH
12.53 °e
17.8 °fH
178 ppm CaCO
3
Chloride content, max.
100 mg/l
Sulfate content, max.
240 mg/l
Carbonic acid content, max.
not detectable
Ammonia content, max.
not detectable
Electrical conductivity, max.
≤ 100 µS/cm
Particle size, max.
< 200 µm
Tbl. 5:
Requirements on the cooling water composition
Installation
23/52