7.5.1 Draining lubricant on the motor side
2
1
Fig. 29:
Draining lubricant on the motor side
1 Magnetic sealing plug/drain screw
2
Collection receptacle
Procedure
1. Place a collection receptacle beneath the drain hole on the motor side.
2. Unscrew the magnetic sealing plug.
3. Fully drain the lubricant.
4. Clean the magnetic sealing plug.
5. Screw in the magnetic sealing plug.
7.5.2 Draining lubricant on the vacuum side
1
2
3
Fig. 30:
Draining lubricant on the vacuum side
1 Magnetic sealing plugs
3
Drain screw
2 Collection receptacle
Maintenance
43/62