2
1
4
Fig. 7:
Exhaust connection with flange connection
1 Exhaust line
3
Centering ring with nozzle and O-rings
2 Protective cap
4
Circlip
Procedure
1. Remove the protective cap from the exhaust flange.
2. Observe the centering ring with nozzle including the corresponding O-rings (2×).
3. Choose a minimum exhaust line diameter equal to the nominal diameter of the connection flange.
4. Depending on the pump type, use PVC or metallic hoses with flange connections from the
5. Route the piping downwards from the vacuum pump, to prevent condensate return.
6. Support or suspend the piping to the vacuum pump so that no piping system forces act on the
vacuum pump.
7. Connect both flanges with a circlip.
5.4 Establishing mains connection
DANGER
Danger to life from electric shock
Contact with exposed and live elements generate an electric shock. Incorrect connection of the mains
supply leads to the risk of live housing parts that can be touched. There is a risk to life.
► Before the installation, check that the connection leads are voltage-free.
► Make sure that electrical installations are only carried out by qualified electricians.
► Provide adequate grounding for the device.
► After connection work, do a PE conductor check.
NOTICE
Risk of property damage from excess voltage
Incorrect or excessive mains voltage will destroy the motor.
► Always observe the motor rating plate specifications.
► Route the mains connection in accordance with locally applicable provisions.
► Always provide a suitable mains fuse to protect the motor and supply cable in the event of a fault.
Installation
24/60
Содержание DUO 1.6
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original DUO 1 6 DUO 1 6 M Rotary vane pump...
Страница 59: ...59 60...
Страница 60: ...PD0061 ed H Date 1907 P N PD0061BEN...