SUPPLEMENTARY INFORMATION

EN

Translation of the Original

XPT 200 PN 

 CCT 36X PN 

 CCT 37X PN

DigiLine gauge with Profinet interface

Содержание CCT 36X PN

Страница 1: ...SUPPLEMENTARY INFORMATION EN Translation of the Original XPT 200 PN CCT 36X PN CCT 37X PN DigiLine gauge with Profinet interface...

Страница 2: ...pfeiffer vacuum de Further operating instructions from Pfeiffer Vacuum can be found in the Download Center on our website Disclaimer of liability These operating instructions describe all models and...

Страница 3: ...1 1 Actual pressure bytes 0 3 21 4 1 2 Actual GCF 1 bytes 4 5 21 4 1 3 Actual GCF 2 bytes 6 7 21 4 1 4 Transmitter status and type byte 8 21 4 1 5 Transmitter warnings and errors byte 9 22 4 1 6 Synt...

Страница 4: ...rofinet master to gauge 23 Tbl 9 Commands and their use 24 Tbl 10 Output data Switching ranges 24 Tbl 11 Behavior and meaning of the Profinet LED 25 Tbl 12 Manufacturer specific diagnosis for CCT gaug...

Страница 5: ...ing and dropping CCT gauge into network area 12 Fig 7 Integration and preparation 13 Fig 8 Integrating input and output modules 14 Fig 9 View after correct configuration 15 Fig 10 Importing GSDML file...

Страница 6: ...Bayard Alpert gauge HPT 200 standard version PG 0024 Operating instructions Digital Pirani cold cathode gauge MPT 200 standard version PG 0025 Declaration of conformity part of these instructions Tbl...

Страница 7: ...point arithmetic for microprocessor systems IO Input output IP Intenet Protocol L A Link activity LED Light emitting diode P Port PN Profinet process field network Profinet Open industrial Ethernet s...

Страница 8: ...ges You can write the correction factors into the memory of the gauge via the serial interface Information can be found in the corresponding operating instructions of the gauge standard version 2 1 1...

Страница 9: ...S 485 address selector switch 3 LED Profinet status on port 1 8 Status LED for the gauge 4 LED communication status 9 RS 485 connector 5 LED Profinet error 2 2 Profinet connection The Profinet connect...

Страница 10: ...peration on both interfaces If you attempt to operate the gauge simultaneously via the RS 485 and Profinet interfaces this will result in incorrect data and interference with the data transmission Ope...

Страница 11: ...ofinet configuration tools There are various suppliers of configuration tools for Profinet configuration The configura tion procedure is identical The figures show the Siemens TIA portal by way of an...

Страница 12: ...Fig 6 Dragging and dropping CCT gauge into network area Installation 12 32...

Страница 13: ...Fig 7 Integration and preparation Installation 13 32...

Страница 14: ...Fig 8 Integrating input and output modules Installation 14 32...

Страница 15: ...CCT 3xx PN there 5 Select the input and output modules for the gauges you are using and then drag and drop all cata log objects from Actual Pressure to Set Correction Factor BA into the active area 6...

Страница 16: ...Fig 10 Importing GSDML file Fig 11 Dragging and dropping the xPT gauge into the network area Installation 16 32...

Страница 17: ...Fig 12 Integration and preparation Installation 17 32...

Страница 18: ...Fig 13 Integrating input and output modules Installation 18 32...

Страница 19: ...ocess control You will find the xPT gauge in the path PROFINET IO Sensors Pfeiffer Vacuum GmbH Digi Line 4 Drag and drop xPT 200 PN into the configuration tool s network area and integrate xPT 200 PN...

Страница 20: ...e following data types are used in the input and output modules Data type Format Float32 32 bit floating point number IEEE Unsigned8 Unsigned 8 bit number Unsigned16 Unsigned 16 bit number Unsigned32...

Страница 21: ...ilable only for gauges with Pirani sensor Bytes 4 and 5 contain the current correction factor 0 2 to 8 0 with 2 decimal places and a factor of 100 for the Pirani sensor Examples Example Correction fac...

Страница 22: ...ge Bit Error Gauge 0 Warning overrange All gauges 1 Warning underrange 2 3 Error Filament 1 defective HPT 200 PN 4 Error Filament 2 defective 5 Error Internal communication All gauges 6 Error EEPROM f...

Страница 23: ...gauge only ever enters the last command executed in Command executed Byte Data type Contents 0 Float32 Adjust Value Pressure 1 2 3 4 Unsigned8 Command 5 Unsigned16 SetData GCF 1 6 7 Unsigned16 SetDat...

Страница 24: ...ion factor 1 00 100 Example Correction factor 8 00 800 4 2 4 SetData GCF 2 bytes 7 8 Bytes 7 and 8 are available only for gauges with HV sensor Bytes 7 and 8 contain the new correction factor 0 2 to 8...

Страница 25: ...ntroller in stop state Bus started synchronization not completed Bus started in configuration mode Profinet device signaling Green flashing 10 Hz Error Green flashing 2 Hz Bus started in configu ratio...

Страница 26: ...rrecting the error disconnecting and recon necting the voltage supply reset User structure identifier 8002 Variable field name Data type Byte Value Error ExtChannelErrorType Unsigned8 0 No error 1 Har...

Страница 27: ...r persists Profinet hardware error 4 1 Switch the gauge off wait 5 seconds and then switch it back on reset 2 Replace the gauge if the error persists Read write error 5 1 Switch the gauge off wait 5 s...

Страница 28: ...max 5 W 15 W Tbl 16 Technical data for Profinet interface with CCT gauges Parameter CPT 200 PN PPT 200 PN RPT 200 PN HPT 200 PN MPT 200 PN Interfaces RS 485 Profinet Profinet interface device side 2 B...

Страница 29: ...ives Low voltage 2014 35 EC Electromagnetic compatibility 2014 30 EU Restriction of the use of certain hazardous substances 2011 65 EU Restriction of the use of certain hazardous substances delegated...

Страница 30: ...Directives Low voltage 2014 35 EC Electromagnetic compatibility 2014 30 EU Restriction of the use of certain hazardous substances 2011 65 EU Restriction of the use of certain hazardous substances dele...

Страница 31: ...31 32...

Страница 32: ...PG0041 ed A Date 2105 P N PG0041BEN...