Characteristics
Units
ACP
28
40
SH-CP-SD
G-LG
CV
SH-CP-SD
G-LG
CV
Power supply
(1)
(according to ordering guide)
V
110-230 V- 1-phase or 200-440 V- 3-phase
Power consumption at ultimate
pressure
W
700
Power consumption at atmos-
pheric pressure
W
1050
1050
1050
1050
1050
1050
Dimensions
(see chapter “Dimensions”, page 49)
Weight
kg
30
32
(1) In accordance with EC regulations, the pumps can withstand a voltage variation of ±10%.
(2) at ambient temperature: 20°C.
(3) Inert gas flushing 300 hPa relative pressure.
(4) Test by Helium spray.
(5) Overall leak test
Tbl. 6:
Technical characteristics
mbar
bar
Pa
hPa
kPa
Torr | mm Hg
mbar
1
1 · 10
-3
100
1
0.1
0.75
bar
1000
1
1 · 10
5
1000
100
750
Pa
0.01
1 · 10
-5
1
0.01
1 · 10
-3
7.5 · 10
-3
hPa
1
1 · 10
-3
100
1
0.1
0.75
kPa
10
0.01
1000
10
1
7.5
Torr | mm Hg
1.33
1.33 · 10
-3
133.32
1.33
0.133
1
1 Pa = 1 N/m
2
Tbl. 7:
Conversion table: Pressure units
mbar l/s
Pa m
3
/s
sccm
Torr l/s
atm cm
3
/s
mbar l/s
1
0.1
59.2
0.75
0.987
Pa m
3
/s
10
1
592
7.5
9.87
sccm
1.69 · 10
-2
1.69 · 10
-3
1
1.27 · 10
-2
1.67 · 10
-2
Torr l/s
1.33
0.133
78.9
1
1.32
atm cm
3
/s
1.01
0.101
59.8
0.76
1
Tbl. 8:
Conversion table: Units for gas throughput
12.2.1 Environmental conditions
Use
indoor use
Installation altitude
up to 2000m
Protection rating
IP20
Ambient operating temperature
12 to 40°C
Storage temperature
-10 to +60°C
Technical data and dimensions
48/56
Содержание ACP 28 CP
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original ACP 28 40 Multi stage Roots pump air cooled...
Страница 54: ......
Страница 56: ...112369 Ed 09 Date 2018 06 P N 112369OEN...