M400 Corpus HD
Troubleshooting - Troubleshooting guide
7 Troubleshooting
7.1
Troubleshooting guide
The following troubleshooting guide describes a number of faults and
events which may occur when you use the wheelchair, together with
suggested remedies. Note that the guide cannot describe all the
problems and events which may occur and you should always contact
your service provider or Permobil in case of doubt.
Event
Possible cause
Remedy
The wheelchair does not start.
Batteries discharged.
Charge the batteries.
The cable connection to the control panel
has come loose.
Insert the cable in the control panel.
Main circuit breaker switched to off
position after e.g. battery replacement.
Reset the main circuit breaker. See
page 73.
Main circuit breaker tripped.
See page 73.
The wheelchair cannot be driven.
Battery charger connected.
Stop charging. Disconnect the charging
cable from the wheelchair charger socket.
Brake release activated.
Reset the brake release.
Wheelchair locked.
Unlock the wheelchair.
The wheelchair switches itself off after a
certain period of inactivity (1 - 30 min).
The electronics’ energy saving mode has
been activated.
Switch the wheelchair on again using the
start button on the control panel.
The wheelchair stops while being driven.
The cable connection to the control panel
has come loose.
Insert the cable in the control panel.
Main circuit breaker tripped.
See page 73.
The wheelchair can only be driven at
reduced speed.
Seat lift or seat angle raised too high.
Applies only to power seat lift and seat
angle.
Lower the seat lift or seat angle.
The wheelchair cannot be charged.
Main circuit breaker switched to off
position after e.g. battery replacement.
See page 73.
The charging circuit breaker has tripped.
Wait five minutes, the circuit breaker will
automatically reset.
7.2
Diagnostics R-net LCD control panel
7.2.1
General
When an error or a fault occurs in the wheelchair’s electronics,
information about it is displayed in the control panel display. This
information can then be used to diagnose where the error, or fault,
occurred and its cause.
Troubleshooting and repairs must always be performed by qualified
personnel with good knowledge of the wheelchair’s electronics.
7.2.2
Diagnostic screens
7.2.2.1
Current diagnostic screen
When the control system’s integrated protection circuits have tripped so
that the control system can no longer operate the wheelchair, a
diagnostic screen is displayed in the control panel display.
86
Содержание M400 Corpus HD
Страница 1: ...Service manual American English M400 Corpus HD ...
Страница 2: ......
Страница 4: ...Text Produced and published by Permobil Edition 3 Date 2018 10 18 Order no 334809 eng US ...
Страница 6: ......
Страница 12: ...M400 Corpus HD Specifications Wiring diagram 3 Specifications 3 1 Wiring diagram 3 1 1 Seat 12 ...
Страница 14: ...M400 Corpus HD Specifications Wiring diagram 3 1 2 Chassis Figure 3 Cabling overview Chassis 1 2 14 ...
Страница 15: ...M400 Corpus HD Specifications Wiring diagram Figure 4 Cabling overview Chassis 2 2 15 ...
Страница 100: ...A 7 330818 350084 334809 eng US ...