Tools menu
110
Using LANCE normalization samples in the assay
Instead of running the LANCE normalization wizard and saving values that can be used for
any subsequent LANCE assay with that particular protocol, you can include LANCE
normalization samples on the plate as part of the assay samples and thus obtain corrected
values for that particular plate.
When you define the protocol make sure that both LANCE labels are defined in the
Measurement tab as shown below. If only one label is selected then the measurement will be
treated as normal time-resolved fluorometry.
When you set up the plate you can select the LANCE well types from the menu that appears
when you click the left mouse button. Use this feature to include at least one of each of the
three LANCE well types on the plate.
You can choose the number of replicates of each well type and you can choose where they
are located on the plate as the following example shows:
Содержание Wallac 1420
Страница 2: ......
Страница 10: ......
Страница 11: ...9 Introduction...
Страница 12: ...10...
Страница 17: ...15 Operating Wallac 1420...
Страница 18: ...16...
Страница 36: ...Operating Wallac 1420 34...
Страница 37: ...35 Protocol editing...
Страница 38: ...36...
Страница 59: ...57 Result viewing...
Страница 60: ...58...
Страница 67: ...65 Tools menu...
Страница 68: ...66...
Страница 128: ......
Страница 129: ...127 Exiting...
Страница 130: ...128...
Страница 132: ...Exiting 130...
Страница 133: ...131 Troubleshooting...
Страница 134: ...132...
Страница 138: ...Troubleshooting 136...
Страница 139: ...137 Appendices...
Страница 140: ...138...
Страница 148: ...Appendix 3 Dispenser operation 146 Preparing the tubing Put the tubing into a container of water...
Страница 165: ...163 Index...
Страница 166: ...164...