Pollution degree
Pollution degree of the intended environment: 2
Transportation and storage
To prevent damage, it is best to store and transport the instrument and its components in their original packing or
other suitable container. Close the packaging with adhesive tape.
Store the instrument in a dry place.
Caution:
Guard the instrument from impact and vibration during transport.
Transportation conditions
0 °C to +60 °C, Relative humidity 10 to 80%.
Storage conditions
0 °C to +60 °C, Relative humidity 10 to 80%.
Safety standards
Certification:
• IEC-CB, CE, NRTL
The instrument fulfills the requirements of:
• IEC 61010-1
• IEC 61010-2-101
• IEC 61010-2-010
• CAN/CSA-C22.2 61010-1
• UL 61010-1
• IEC 61326-1
• IEC 61326-2-6
Optional features and spare parts
The following optional features can be ordered separately from PerkinElmer:
•
Cooling block
: the cooling block is designed to cool the plates once the incubation/shaking is over. The cooling
block allows for quicker cooling of the nests. Typically, one cooling block in the middle nest will do but for rapid
cooling, place one cooling block in each nest. Note that the cooling block's own temperature affects the cooling
speed.
Caution:
Do not use cooling blocks while the instrument is shaking. There is no locking mechanism for cooling blocks
and their movement, produced by the shaking, will cause damage to the inner coating of the chamber.
TriNEST Microplate incubator and shaker
Specifications and features
21