90 . Spotlight 200 User’s Guide
Cleaning the ATR Objective Crystal
Because the ATR objective crystal is in contact with the sample under test, it may become
dirty during use. A dirty crystal may give spectra of the contaminant rather than the sample
under test.
Great care must be taken when cleaning the crystal. It should be cleaned using the minimum
of mechanical pressure, with a soft brush or lens tissue. Isopropanol or n-hexane may be
used to clean the crystal.
NOTE:
Do not use acetone or xylene to clean the crystal.
1.
Pour a small volume of a solvent into a shallow dish.
2.
Using the
Z
-control on the joystick, lower the sample stage, to allow the dish to be
placed on the stage under the crystal.
3.
Lower the crystal.
4.
Using the
Z
-control on the joystick, raise the stage, so that the tip of the crystal is just
immersed in the solvent.
NOTE:
Do not immerse the whole of the ATR crystal in the solvent.
5.
Leave the crystal in the solvent for 5 minutes.
We recommend that you do not leave the crystal in the solvent for longer periods.
6.
If any material remains on the crystal, rub the crystal gently with a soft brush or lens
tissue.
Содержание Spotlight 200
Страница 1: ...Spotlight 200 User s Guide MOLECULAR SPECTROSCOPY...
Страница 6: ...6 Spotlight 200 User s Guide...
Страница 7: ...Introduction...
Страница 13: ...Warnings and Safety Information...
Страница 28: ...28 Spotlight 200 User s Guide...
Страница 29: ...Overview of the Spotlight 200...
Страница 40: ...40 Spotlight 200 User s Guide...
Страница 41: ...Getting Ready to Use the Spotlight 200...
Страница 51: ...Preparing Samples...
Страница 65: ...Techniques for Collecting Spectra...
Страница 112: ...Maintenance...
Страница 126: ...126 Spotlight 200 User s Guide...
Страница 127: ...Appendices...