68 . Raman IdentiCheck Getting Started Guide
Cleaning the probe body and cable
Wipe all external surfaces with a lint-free cloth. If necessary, this cloth may be dampened
with a mild detergent solution.
Replacing the Triggered Fiber Optic Probe
The TFOP is constructed such that a single armored cable bifurcates into an excitation
cable and a collection cable. The excitation fiber transmits the laser energy, whereas the
collection cable is a conduit for the Raman energy. Both fibers terminate in FC style
connectors.
Removing a probe
1.
Ensure the attenuator cap is attached to the probe.
2.
Disconnect the instrument from the power supply.
3.
Unscrew and remove the probe cable retaining bar (Figure 17).
Figure 17 Probe connectors
CAUTION
Do not use copious amounts of water or allow the water to enter the
probe casing.
Probe cable
retaining bar
Probe
couplers
Retained
screw-in caps
Содержание Raman IdentiCheck
Страница 1: ...Raman IdentiCheck Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Raman IdentiCheck Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 26: ...26 Raman IdentiCheck Getting Started Guide ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 Raman IdentiCheck Getting Started Guide ...
Страница 35: ...Introduction to the Raman IdentiCheck ...
Страница 50: ...50 Raman IdentiCheck Getting Started Guide ...
Страница 51: ...Using the Raman IdentiCheck ...
Страница 65: ...Maintenance and Troubleshooting ...
Страница 73: ...Appendix A Using your own PC ...
Страница 78: ...78 Raman IdentiCheck Getting Started Guide ...