Maintenance and Troubleshooting . 71
Troubleshooting
This section is a reference guide if you are having any problems with your instrument.
Troubleshooting outlines what to do if your instrument or software malfunctions.
Hardware
Problem:
During the installation procedure the USB devices are not
recognized.
Description This may mean that there is not enough power on the USB sockets on your
PC.
Solution
Ensure that your PC has USB 2.0 sockets. If you are not sure, use a USB 2.0
power hub.
Problem:
Cannot collect data because an interlock is breached.
Description This means that, for some reason, the interlock system is detecting an open
circuit.
Solution 1
Ensure all interlocks, including any remotely interlocked doors, are closed
properly.
Solution 2
Ensure your external interlock is wired up correctly and that any interlocked
room entrances or sampling enclosures are closed.
Problem:
You are prompted for the CCD drivers.
Description This may happen if you disconnect the instrument control USB cable and
reattach it to a different USB port, or if you switch the instrument off and
then switch it on again.
Solution
Refer to the
Spectrum Administrator’s Guide
(part number L1050095) for
details of how to reinstall the drivers.
Software
Problem
Spectrum software stops responding.
Description When you are using the software, for some reason it stops working; that is,
mouse clicks have no effect.
Solution
Hold down the CTRL+ALT+DELETE keyboard buttons to display the
Windows Security dialog. Click
Task Manager
and then click the
Processes
tab. In the list under
Image Name
look for
IRWinLab.exe
(to find this you
may have scroll through the list using the scroll bar on the right side) click
on this, and then click
End Process
.
Then select
Raman.exe
and then click
End Process
.
If a
Task Manager Warning
dialog is displayed click
OK
. Re-
start the software. If you are unsure how to do this please refer to SOP-1.
If you do not intend to restart the software, you must switch the
power to the instrument off. Otherwise the laser may not shut
down.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...