Using the RamanFlex 400 Series . 63
Summary
•
The only calibration procedure that needs to be performed on a regular basis is the
wavelength calibration verification procedure.
•
The wavelength and laser frequency calibrations should only be performed if the
instrument fails the wavelength calibration verification procedure.
•
The intensity calibration should only be performed on a yearly basis, or if the
environment (temperature and / or humidity) changes significantly, or if the instrument
is moved, or following a service repair.