Using the RamanFlex 400 Series . 61
8.
Select
.
The Frequency Calibration dialog is displayed.
9.
Ensure that you have followed the instructions on the dialog, and then click
Next
.
The instrument will now perform the laser frequency calibration routine. This will take a
few minutes. At the end of the calibration, the result is displayed.
In the unlikely event that the laser frequency calibration routine should fail, please
perform a wavelength calibration and repeat the laser frequency calibration.
NOTE: If the calibration fails again, please contact your PerkinElmer Service Representative.
SOP-10: Intensity Calibration
The intensity calibration must be performed using the National Institute of Standards and
Technology (NIST) Standard Reference Material. NIST SRM 2241 is the relative intensity
correction standard for Raman spectroscopy when using 785 nm excitation.
NOTE: NIST calibration is temperature dependent. If you intend to perform quantitative
analyses, you should ensure that your instrument is NIST calibrated at the
temperature at which data is to be acquired. Otherwise, any data should be used for
qualitative analyses only.
The NIST intensity correction is valid in the recommended operating temperature
range 20
°
C – 25
°
C. If you move the instrument to an environment outside this
temperature range, that is 15
°
C – 20
°
C or 25
°
C – 30
°
C, we recommend that you
perform a NIST intensity correction at this new temperature.
Contact your PerkinElmer Raman Business Unit if you wish to set up a NIST intensity
correction for a different temperature range.
Intensity calibration should typically be performed on an annual basis – that is, during an
annual instrument service – but also whenever the instrument is moved, following a service
repair, or when there has been significant temperature variation in the local environment.
NOTE: As intensity calibration is not required on a regular basis, NIST Standards are not
included in the basic RamanFlex 400 package. Standard Reference Materials can be
purchased from PerkinElmer (part number L1321831). Your instrument will be
intensity calibrated prior to shipping or during installation.
NOTE: If you have a RamanFlex 400F with two fiber optic probes, you must perform an
intensity calibration on both fiber optic probes.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...