Operation . 41
Local Control Mode
Starting Temperature Control
NOTE:
As previously mentioned, it is good practice to allow the system to reach a stable
temperature before inserting a sample. Once the temperature is stable, this control
can be useful for pre-heating samples before running a cycle.
1.
Press or hold FUNC until
C-01
,
C-02
,
S.SP
or
Conf
is displayed in the Lower display
(STOP status).
2.
Use the arrow buttons
and
to scroll through these modes until
S.SP
is
displayed in the Lower display.
The Upper display shows the current temperature in the cell changer.
3.
Press FUNC.
The temperature control starts. The Lower display shows the set temperature and the
Upper display shows the actual temperature.
Modifying the Setpoint
1.
When standard temp control is running press the arrow buttons to change the setpoint.
2.
To save the setpoint release the arrow buttons.
Stopping Temperature Control
¾
Hold FUNC until the system ‘beeps’.
The system returns to STOP status and maintains this setpoint.