38
Preparing Your Laboratory
Lab Space Requirements
The work bench must be sufficiently sturdy to carry the weight of the PinAAcle
900 Series, computer with monitor, keyboard, mouse and printer. The weight of the
optical accessories must also be taken into account. The maximum load imposed
by the instrument is 0.5kN/m
2
. Check that the floor can withstand this load.
At least 15 cm (6 in) free space must be available behind the instrument for
ventilation purposes and servicing. Do not block the fans operating on both sides
of the instrument.
In addition, an area of at least 69 cm (27 in) should be available. If this is not
possible, the bench on which the instrument is installed must be on wheels that can
be locked.
Make sure that there is enough room on the work bench. The monitor can usually
be placed on top of the PC unit in order to save working space. Some clearance
between the instruments should be provided to ensure sufficient air circulation.
If a printer is included in the delivery, additional space must be provided. Do not
forget the space required for a mouse and mouse pad.
Optional Accessories (e.g., AutoPrep 50) require additional space. It is
recommended to install these accessories on the sample tray or on a sample cart in
front of the instrument.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......