153
Installing the Sample Capillary
Checking the solution droplet delivery
Once you have aligned the autosampler as described in the preceding sections, you
can check that solutions will be correctly dispensed by injecting blank solution into
the graphite tube.
1. Create a furnace program comprising drying, pyrolysis, atomization, and
clean-out.
Enter the total volume of solution that you will use in the subsequent analyses.
Refer to the
Online Help
in WinLab32 for AA on how to create a method.
2. Place a sample cup containing blank solution in location 1 in the autosampler.
Use an identical blank solution to the one that you will use for the subsequent
analyses.
3. In the
Automated Analysis Control
window, on the
Setup
page, in the
Locations
column, enter the autosampler location of the blank solution.
4. Start the method:
In the
Automated Analysis Control
window, on the
Analyze
page, click on the
Analyze All
button. The
Analysis Warning
screen will appear click on
OK
.
5. Look in the TubeView camera window and watch the delivery of the solution
droplet onto the platform.
6. If the solution droplet delivery was unsatisfactory, make the appropriate small
adjustments to the penetration depth of the pipette tip. See
fine-adjustments to the penetration depth of the pipette tip in the graphite tube
on page 150
.
7. Repeat step 4 through step 8 to dispense further aliquots of blank solution until
you are satisfied with the delivery.
Warning: UV Radiation – Risk of Eye Damage
The lamps may emit UV radiation which can damage your eyes.
Always wear UV-absorbing safety glasses when aligning the
autosampler.
!
Warning
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......