Instrument overview
The Migele Gel Electrophoresis Unit is used together with the programmable power supply (CWP-2000 or equivalent)
and the circulating water bath (1023-0010, 1023-0020) or equivalent. The instrument can run gels of different sizes. For
example, 240 × 203 mm (FR-9400, FR-9360) and 203 × 127 mm (FR-9120).
Additional equipment required but not supplied with the instrument:
• Programmable power supply CWP-2000 or equivalent
Voltage at least 1500 V
Power at least 80 W
Current at least 200 mA
• Circulating water bath (1023-0010, 1023-0020) or equivalent
Temperature s10 - +15 °C
Flow rate at least 5 l/min
Maximum pressure below 0.5 bar
Cooling power at least 160 W at +10 - +15 °C
Figure 1: Migele Gel Electrophoresis Unit
2118-0010 - Migele
®