Using Sample Names Files
60
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Using a Barcode to Specify the Sample Names
File (Continued)
6
Touch the
Start Run
button. The
opens.
7
Touch the
Export
button and save the Run file. The Run File
can be imported to run multiple plates with the same settings.
8
To start the run, touch the
Start
button.
9
When the run is started, the plate barcode is read and the
Sample Names file with the corresponding barcode is used to
select the wells to sample. In the example above, if the plate
barcode is 12345, then the file named “Sample
Names12345.csv” will be used to select the wells in the run.
Editing Sample Names During a Run
The default Sample Names assigned during a run can be edited
after the run starts, but before the sample well read is complete.
(After the well has been read, the data is saved in the data file, and
the sample name can no longer be changed.) same names
assigned using a sample names file can also be edited before the
well has been read.
1
After a run is started, touch the
Views
button on the
2
Touch the
Sample Names
button to open the
. If a sample names file was specified, the names from the
Sample Names File display. Otherwise, the well labels are used
as the default sample names.
3
Edit the sample names as desired. (If a well has already been
read, the Sample Names and Comment fields are disabled for
the well and cannot be edited.)
4
Add comments to the Comment fields as desired.
5
Touch the
Apply
button to save the changes. You must touch
Apply before the well read is complete.