Sodium Bullet Test
•
Precision Test:
This test indicates the Relative Standard Deviation (RSD) of
the instrument's analyte emission intensity or concentration measurements.
The RSD may also be referred to as the CV (Coefficient of Variation). A high
RSD or CV is usually indicative of a problem with the sample introduction
system such as improper drainage, leaks, improper tension on the pump tubing,
worn pump tubing, or nebulizer problems. Refer to the Syngistix Software
Guide on the Syngistix Document Pack CD for details
.
•
Detection Limit Test:
This test measures the baseline signal for a blank
solution to give an indication of the lowest concentration of an element which
can be measured. The detection limit is calculated as three times the standard
deviation of the blank. Refer to the Syngistix Software Guide on the Syngistix
Document Pack CD for details
.
Sodium Bullet Test
To run the sodium bullet test:
1. For aqueous analyses, aspirate a 1000 mg/L solution of sodium. For organic
analyses, aspirate an appropriate organic blank.
2. Examine the plasma through the viewing window in the sample compartment
door.
A yellow-orange bullet (or green, in the case of organic compounds) should be
visible in the center of the discharge, and should extend from the base of the
discharge to just below the top plasma induction plate. If the bullet height is
unsatisfactory, adjust the nebulizer argon flow in the Method Editor or Plasma
Control windows.
If no bullet appears or the bullet is faint:
•
Check that your sample does contain sodium at the required concentration.
•
Check that sample is being pumped to the nebulizer. Make sure that the pump
lever is engaged and that the pump tubes are connected appropriately.
•
Check that the drain is being pumped properly. You should see a segmented
flow of bubbles through the tubing.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......