291
Peristaltic Pump
Installing the Drain Tubing
Install the drain tubing on an open channel of the peristaltic pump as follows:
1. Pre-stretch a piece of 1.14 mm (0.045 in.) pump drain tubing.
2. Carefully stretch the new tubing around the pump head and place the clips on
the pump tubing into the slots, using the slots in the same position on each side
of the pump head. Be careful not to twist or kink the pump tubing during this
process.
3. Attach the Teflon tubing to the pump tubing for the drain.
Make sure you install the tubing on the pump so that the waste liquid is
pumped
out
of the spray chamber. See Figure 5-40.
If the tubing is installed
in the wrong direction, the spray chamber will flood.
4. For the tubing that leads to the drain bottle, use one piece of 1.07 mm i.d.
(0.042 in.) Teflon tubing (Part No. 09985708). Bevel one end by cutting it at
an angle.
5. Insert the Teflon tubing into the other side of the pump tubing. Lead this tubing
to the drain bottle. Refer to Figure 5-40.
6. Adjust the tubing as described in the next procedure.
3
Sleeve (Part No. N0695415)
8
Teflon Tubing 1 mm i.d.
(Part No. 09985708)
4
Spray Chamber
9
Drain bottle
5
PVC Tubing 1.5 mm i.d.
(Part No. 09985735)
10
Peristaltic Pump
Item
Description
Item
Description
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......