Troubleshooting
287
Autosampler Problems
Table 6-4.
Autosampler Problems
Problem
Possible Cause(s)
Corrective Action
Autosampler does not
respond
There may be a
communication problem.
Use the
Instrument Diagnostics
window to reset the autosampler.
The software may not be
configured for the autosampler
you are using.
Select the
Instrument tab -> Devices
click on the
Autosampler
tab and check
the autosampler configuration. If the
wrong autosampler model is shown exit,
Syngistix and run the
Syngistix
Reconfiguration
program from the
Windows start menu to change the
autosampler model.
One of the cables may have
been disconnected or the
autosampler may have been
turned off.
Check the power cable and the cable
connecting the autosampler to the
computer. Make sure the power is on.
For AS-93plus and other GPIB
controlled autosamplers, the
DIP switches may be set
incorrectly.
Reference the appropriate autosampler
hardware guide for the correct DIP
switches.
The pins on the autosampler
cable connector may be
broken.
Check the connector. If the pins are
broken, order a replacement cable.
Autosampler does not
go to the correct
location.
The software may not be
configured for the correct
autosampler tray.
Autosampler tray is selected by going to
Instrument
tab and selecting
Devices
and then select the
Autosampler
tab. In
the
Autosampler Setting
field in the
Autosampler
tab select the tray type
from the browse window.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......