Troubleshooting
279
Performance Problems: Troubleshooting Table
Table 6-1.
Performance Problems: Troubleshooting Table
Problem
Possible Cause(s)
Corrective Action
RSD (Relative
Standard Deviation)
greater than 1%.
The sample introduction
system may not be set up
correctly or may require
maintenance.
Refer to
Does the sample have a high
viscosity or high percentage of
dissolved solids?
Try increasing the integration time.
Refer to the Syngistix Software Guide
on the Syngistix Document Pack CD
for details
.
For high dissolved solids, try the
GemCone spray nebulizer.
Add 0.05% surfactant to your samples,
standards, and blanks.
Try running other samples with normal
characteristics to determine if the
sample matrix is really the problem.
The standard and blank may need to be
matrix-matched. Or, internal
standardization may be required.
Look for carryover – check if
concentrations of replicates
increase from one to the next.
Increase the Read Delay to flush the
sample introduction system completely.
Increase the rinse time if using an
autosampler. Prepare the rinse solution
in the same acid matrix used for the
standards, samples, and blank. Refer to
the Syngistix Software Guide on the
Syngistix Document Pack CD for
details
.
RF power and/or nebulizer
argon flow may require
adjustment.
Adjust as necessary. Also, check that
you are using the recommended
settings for your instrument.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......