Maintenance
200
Removing and Replacing the Windows
Normally you should remove the axial window before you remove the radial
window. This releases the pressure in the spectrometer and helps avoid objects or
contamination falling into the radial snout.
Removing and Returning the Axial Purge Window
Tool Needed:
•
Purge window aluminum
ring removal tool (Part No. N0790448)
1. Turn off the plasma.
2. In order to gain access to the Axial Purge window, you must remove the Shear
Gas Nozzle. See
Caution
Attention
These windows are important parts of the optical system. Handle them
carefully, as you would any sensitive optical component. Do
not
directly
touch the windows. Always remove the axial window before the radial
window and refit the radial window before the axial window. This will
help avoid objects or contamination falling into the radial snout.
Ces fenêtres sont des éléments importants du système optique. Manipulez-
les avec soin, comme vous le feriez pour tout composant optique sensible.
Ne touchez pas directement les fenêtres. Retirez toujours la fenêtre axiale
avant que la fenêtre radiale et remonter la fenêtre radiale devant la fenêtre
axiale. Cela permettra d'éviter des objets ou la contamination de tomber
dans le museau radial.
Caution
Attention
Wear
cotton gloves
when you handle the torch glassware
.
Porter des gants de coton lorsque vous manipulez la verrerie de la torche.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......