Maintenance
193
manufacturer's safety recommendations. These recommendations are normally
provided in a material safety data sheet (MSDS) supplied with the chemical.
If you have been analyzing organic solutions, the spray chamber can be cleaned
using a solvent or diluted soap solution. Be sure to inspect the torch and injector daily
for carbon buildup.
O-rings may be cleaned with soap and water. An ultrasonic bath may be used.
Replace if cracked or worn. A torch O-ring kit is available. See the list of
replacement parts at the end of this chapter.
Replacing the Scott-Type Spray Chamber
See the procedure
Connecting the Scott Spray Chamber
Cyclonic Spray Chamber
The cyclonic spray chamber is a glass spray chamber. The cyclonic spray chamber
may be used with the concentric glass, low-flow GemCone, and high flow GemCone
nebulizers.
There are two cyclonic spray chambers. The baffled cyclonic spray chamber (Part
No. N0791352) is recommended for all aqueous and most organic solutions. For
really volatile organics, cyclonic spray chamber (Part No. N0777497) can be used.
Removing the Cyclonic Spray Chamber
1. Disconnect the tubing to the cyclonic spray chamber.
Handle the cyclonic spray chamber carefully to avoid breaking the
glassware and possible injury.
Avertissement: Manipuler avec soin la chambre de nébulisation
cyclonique pour éviter de casser la verrerie et des blessures.
!
Warning
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......