
68153 01
www.perkinelmer.com
39
XRpad2 4336
D
IGIT
AL
I
MAGING
Table 15
Guidance and Manufacturer’s Declaration of Electromagnetic Immunity
(Portable Equipment)
Guidance and Manufacturer’s Declaration of Electromagnetic Immunity
The X-ray detector is intended for use in the electromagnetic environment specified below. The installer,
X-ray system manufacturer, or user of the X-ray detector should assure that it is used in such an
environment.
Immunity Test IEC 60601 Test
Compliance
Electromagnetic Environment – Guidance
Conducted
radio-frequency
fields (CEF)
IEC 61000-4-6
Radiated
electromagnetic
field (REF)
IEC 61000-4-3
3 V
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
[V1] 3 V
150 kHz to
80 MHz
[E1] 3 V/m
80 MHz to
2.5 GHz
Portable and mobile RF-communication equipment
should not be closer to any part of the X-ray
detector including the data cables, than the
recommended separation distance calculated from
the equation appropriate for the frequency of the
transmitter.
, for 150 kHz to 80 MHz,
, for 80 MHz to 800 MHz,
, for 800 MHz to 2.5 GHz,
where
P
is the maximum output of the transmitter
in watt (W) according to the transmitter
manufacturer and
d
is the recommended
separation distance in meter (m).
Field strengths outside the shielded location from
fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than
3 V/m.
Interference may occur in the vicinity of equipment
marked with the following symbol.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the X-ray detector is used exceeds the applicable RF compliance level above, the
X-ray detector should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the X-ray detector.
Note 1:
These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Note 2:
It is essential that the actual shielding effectiveness and filter attenuation of the shielded location
be verified to assure that they meet the minimum specification.
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2
Содержание XRpad2 4336
Страница 14: ...This page intentionally left blank xiv www perkinelmer com 68153 01 List of Figures XRpad2 4336 ...
Страница 16: ...xvi www perkinelmer com 68153 01 List of Tables XRpad2 4336 This page intentionally left blank ...
Страница 62: ...This page intentionally left blank XRpad2 4336 46 68153 01 ...
Страница 66: ...This page intentionally left blank 50 www perkinelmer com 68153 01 Index XRpad2 4336 ...