68153 01
www.perkinelmer.com
37
XRpad2 4336
D
IGIT
AL
I
MAGING
Table 13
Guidance and Manufacturer’s Declaration of Electromagnetic Immunity
Guidance and Manufacturer’s Declaration of Electromagnetic Immunity
The X-ray detector is intended for use in the electromagnetic environment specified below. The installer,
X-ray system manufacturer, or user of the X-ray detector is responsible for the usage condition of the
detector to be within such environment.
Immunity Test
IEC 60601 Test
Compliance
Electromagnetic Environment –
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Contact: 6 kV
Air: 8 kV
Contact: 6 kV
Air: 8 kV
Floors should be made of wood, concrete,
or ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transients (Burst)
IEC 61000-4-4
0.5 kV (AC)
1 kV (DC)
0.5 kV (AC)
1 kV (DC)
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Transients-Surges
IEC 61000-4-5
1 kV /‘2 kV
1 kV /2 kV
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Power frequency
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial and/or hospital
environment.
Voltage dips and
short interruptions
IEC 61000-4-11
–95%/10 ms
–60%/100 ms
–30%/500 ms
>–95%/5000 ms
–95%/10 ms
–60%/100 ms
–30%/500 ms
>–95%/5000 ms
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the X-ray detector requires
continued operation during power mains
interruptions, we recommend that the
X-ray detector be powered from an
uninterruptible power supply or battery.
Содержание XRpad2 4336
Страница 14: ...This page intentionally left blank xiv www perkinelmer com 68153 01 List of Figures XRpad2 4336 ...
Страница 16: ...xvi www perkinelmer com 68153 01 List of Tables XRpad2 4336 This page intentionally left blank ...
Страница 62: ...This page intentionally left blank XRpad2 4336 46 68153 01 ...
Страница 66: ...This page intentionally left blank 50 www perkinelmer com 68153 01 Index XRpad2 4336 ...