21
XRpad2 4336
D
IGIT
AL
I
MAGING
8.5.2.1
Cleaning the XRpad IPU-2
If the XRpad IPU-2 (Interface and Power Unit) surface is dirty or dusty, clean it with a cleaning
cloth dampened with ethanol or a diluted neutral detergent. If you are using a disinfectant other
than those specified, we recommend you consult a specialist for the procedure for disinfection.
Turn off the XRpad IPU-2, and disconnect the AC power cable, detector power, and detector
communication tethered cables before cleaning.
8.6
Minimum System Requirements
The following are the minimum requirements for the host computer that controls the X-ray
detector.
1
Gigabit Ethernet Infrastructure and a free Gigabit Ethernet Port or Wi-Fi Infrastructure.
2
Intel compatible Multi Core Processor (> 2 GHz).
3
RAM > 4 GB.
4
Windows 7 (32-bit/64-bit).
5
If a Firewall is used, make sure that it allows a connection to the detector.
6
Access Point:
a
WPA2 encryption support.
b
802.11n standard with 20 MHz channel bandwidth 36-48, 52-64, 100-116, 132-140,
149-165.
c
MIMO 3x3.
d
Complying with IEC 60601-1 or IEC 60950-1.
WARNING
When the Interface Power Unit is going to be cleaned, be sure to turn
off the XRpad IPU-2 (Interface and Power Unit), and unplug all cables.
Never use thinner, benzine, acetone, or other flammable cleaning
agents. Ignoring this warning may result in explosion, fire, or electric
shock, which may result in severe personal injury, death, or
substantial product damage.
Содержание XRpad2 4336
Страница 14: ...This page intentionally left blank xiv www perkinelmer com 68153 01 List of Figures XRpad2 4336 ...
Страница 16: ...xvi www perkinelmer com 68153 01 List of Tables XRpad2 4336 This page intentionally left blank ...
Страница 62: ...This page intentionally left blank XRpad2 4336 46 68153 01 ...
Страница 66: ...This page intentionally left blank 50 www perkinelmer com 68153 01 Index XRpad2 4336 ...