![Perkin-Elmer RamanStation 400 Series Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/perkin-elmer/ramanstation-400-series/ramanstation-400-series_getting-started-manual_1530158094.webp)
94 . RamanStation 400 Series Getting Started Guide
Replacing the Stage Controller Fuse
NOTE: This section is only applicable if you have an motorized XYZ stage.
WARNING
Electrical hazard
To prevent potential injury to yourself and damage to the instrument,
switch OFF all components in the system and disconnect them from the
line power supply before you alter, or make any new, electrical
connections.
Removal of the protective housing or exterior panels on any instrument
will allow potentially hazardous Class 3B laser radiation to be emitted
and will invalidate the instrument warranty.
1.
Switch off the stage controller, disconnect it from the power supply and remove the
mains cord.
2.
Insert a screwdriver into the slot at the side of the fuse drawer, and pull out and flip to
one side over the mains inlet.
The fuse may now be removed.
3.
Fit the replacement fuse into the fuse drawer.
Make sure that you fit the fuse in the top slot. You require a 1.6 A time-lag fuse
(09991641). Always use the recommended replacement fuse.
4.
Refit the fuse drawer.
The only other fuses that you may have to replace are those in the plugs of the mains power
cords for the spectrometer and microscope (UK-type plugs only).
Содержание RamanStation 400 Series
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanStation 400 Series Getting Started Guide ...
Страница 35: ...Preparing Your Facility ...
Страница 39: ...Installing Your RamanStation 400 Series ...
Страница 52: ...52 RamanStation 400 Series Getting Started Guide ...
Страница 53: ...Introduction to the RamanStation 400 Series ...
Страница 59: ...Fiber Optic Probe ...
Страница 66: ...66 RamanStation 400 Series Getting Started Guide ...
Страница 67: ...Using the RamanStation 400 ...
Страница 91: ...Maintenance and Troubleshooting ...